8.7. Bluetooth Measurements Not Displayed/Refreshed
If continuous acquisition is activated, measurement data appears the Measurements menu (
). If
measurements do not appear as expected:
Ensure the device is powered ON and is in the desired measurement mode, and that data acquisition on
the camera has not been stopped (in other words, the image is frozen). To preserve consistency between
the image displayed and the Bluetooth measurements read on the devices at the time of acquisition, the
measurements are refreshed only when there is a new acquisition. If a device is connected, but
acquisition has been stopped, its measurements will not be displayed until there has been another
acquisition.
Following an incorrect reading, the measurement read and the measurement expected may not be
consistent. This can prevent refreshing. To eliminate this, change the measurement function on the
device and then change it back to the original function.
If the problem persists:
1. Switch OFF the Bluetooth peripheral.
2. Wait until this disconnection is detected on the camera.
3. Turn the peripheral back ON (with Bluetooth mode activated).
4. Go to the Peripherals menu (
) and press
F2
to attempt reconnection.