6. Connect one end of the leads to the
Ω
and
COM
terminals and
the other end to the sample under test.
The measurement appears on the LCD.
The instrument displays the measured value minus the
compensation. The value displayed may be negative if the
leads used for the measurement are not the ones that were
compensated. In this case, repeat the compensation. Lead
compensation is preserved after Auto Off but not after the
instrument is manually turned OFF.
NOTE: Error messages:
> 99.99 kΩ
: The measured value falls outside the instrument’s
measurement range.
NOISE
: A stray or foreign voltage U
HE
> 3 V is detected between the
Ω
and
COM
terminals.
: : A stray or foreign voltage U
HE
> 50 V is detected between the
Ω
and
COM
terminals; in this case measurement is not possible.
NOTE:
That
2P
mode is useful for ensuring the H electrode
connection is intact.