20
Ground Resistance Testers Models 6422 & 6424
2.2.2 Model 6424
NOTE:
The sample under test should be de-energized.
1. Turn ON the instrument by pressing
for > 2 seconds. The
symbol
2P
appears on the right side of the LCD.
2. If the Model 6424 is already ON but in voltage or current
measurement mode, press
to access the resistance
measurement mode.
3. Before making measurements, perform lead compensation. This
subtracts the test lead resistance from the measurement. Connect
one end of the leads to the
Ω
and
COM
terminals and touch the
other ends together, creating a short-circuit.
4. Press the
button.
5. Press
for > 2 seconds. The
symbol blinks during the
compensation procedure. When finished, the LCD displays
00.00Ω
.
NOTE:
If the message
Err
appears, the compensation procedure
failed, either because the value to be compensated exceeds 5 Ω or
because the leads were disconnected during compensation.
Press
to deactivate/reactivate lead compensation.