Megohmmeter Model 1015
21
WARNING:
Make sure that no terminals are connected and that
the switch is in the
OFF
position before opening the back of the
instrument.
• Remove the yellow shockproof housing.
• Remove the two screws from the back of the instrument (Figure 11).
• Remove the back cover.
• Replace the fuse or batteries.
NOTE:
The instrument will operate correctly with a battery voltage
range of (4.5 to 6.5) V.
• Reverse the above procedure to install the back cover and the shock proof
housing.
•
To close the back stand flush with the back cover, apply light pressure at
Point A in Figure 11.
Figure 11
5.2 CLEANING
WARNING:
Before cleaning, disconnect all inputs to prevent
possible electric shock.
The instrument should be cleaned periodically to keep the LCD clear and
prevent the buildup of dirt and grease around the instrument’s buttons.
Clean the body of the instrument with a damp and soapy cloth
Do not submerge the instrument in water
Do not use solvents
+
+
+
+
Cover screws
1.5 V batteries LR6
1.6 A fuse
Point A