Megohmmeter Model 1015
11
3.2 GENERAL SPECIFICATIONS
Power Supply:
Four 1.5 V AA batteries
Dielectric Test:
6000 V, 50/60 Hz, 1 m
Meter Movement:
Rugged taut band suspension
Dimensions:
(6.6 x 4.2 x 2.2) in (167 x 106 x 55) mm
Weight:
1.45 lb (650 g)
Operating Temperature Range:
(14 to 131) °F (-10 to 55) °C, (20 to 80) % RH
Storage Range:
(-40 to 158) °F (-40 to 70) °C, (10 to 90) % RH
Terminals:
“-” Line (black); “+” earth (red); accepts 4 mm banana plugs
Reference Conditions: (63 to 73) °F (17 to 23) °C, (45 to 55) % RH; Supply Voltage:
(5.3 to 5.7) V; Voltage Frequency: (45 to 65) Hz; Electric Field: < 1 V/M; Magnetic
Field: < 40 A/M; Position: Horizontal ± 5°.
3.3 SAFETY SPECIFICATIONS
IEC 1010-1:
Safety requirements for electrical equipment for measure-
ment, control and laboratory use
VDE 0413-1:
Insulation testers
VDE 0413-4:
Ohmmeters
VDE 0100:
Specifications for the creation of high current installations of
nominal voltage less than 1000V
IEC 801:
Electromagnetic compatibility for measurement and control
equipment in industrial processes
Part 2: Sections relating to electrostatic discharges
Part 3: Sections relating to radiated electric fields
Part 4: Sections relating to electrostatic discharges
Part 5: Sections relating to electric shocks
IEC 68.2.6:
Vibrations
IEC 68.2.27:
Shocks
IEC 68.2.29:
Shakes
IEC 68.2.31:
Bumps
IEC 68.2.32:
Free fall
UL 94:
Self-extinguishing capability
Specifications are subject to change at any time without notice