Press  UP  until  INT  1  is  highlighted,  then  press  ENTER  and  EN  (enabled) 
appears.  Repeat the above until INT 2 is also enabled.  If you are only using 
one RF Interlock, make sure that at least one of the Interlocks is enabled.   If 
only one Interlock is enabled, say INT 1, press DOWN.  INT  2 will appear as 
DIS.    OK  will  then  appear,  highlighted  in  blue.    Press  ENTER.    NEXT  will 
appear  highlighted  in  blue.    Press  ENTER.    Laser  Standby!  will  appear  in 
green.  CONT will be highlighted in blue, press ENTER and an image of laser 
safety eyewear appears along with the words ‘Safety glasses must be worn’ 
appearing in red.  The next screen states: 
 

 
 
 
 

 
 
 
 
 

START will appear in red.   
 
It is now advisable to have the laser ready at the position where the subject is 
to  receive  treatment.    Press  the  red  START  /  STOP  button  and  the  AEL  IR 
4400 delivers the requested dosage.  An audible beep  will be heard and the 
Laser  Emission  indicator  button  on  the  front  of  the  unit  will  be  flashing  in 
green.    A  timer  will  appear  indicating  the  amount  of  time  left  for  treatment.  
The  AEL  IR  4400  will  countdown  the  treatment  time.  The  minutes  and 
seconds left will appear in orange text on the screen.  The words Laser ON! 
will be in red on the screen.   

 
Repeating a dosage: 

Upon completion of the treatment, a long beep will be heard, followed by the 
words  Laser  Standby!  in  green  on  the  screen.  NEXT  will  be  highlighted  in 
blue.      Press  ENTER,  an  image  of  laser  safety  eyewear  appears  along  with 
the  words 

‘Safety  glasses  must  be  worn’  in  red.    Then  press  the  START  / 

STOP  button.    START  will  be  highlighted  in  red.    The  same  dosage  will  be 
delivered. 
 

 
 
 
 

 
 

Summary of Contents for IR 4400

Page 1: ...AEL IR 4400 Therapy Unit User Manual...

Page 2: ...1...

Page 3: ...7 Indicators and Controls 8 Caution Statement 8 Operating Instructions 8 9 Operating the unit 9 17 Care and Maintenance 18 Warnings 18 Testing the RF Battery 18 Specifications 19 Battery Re charging 2...

Page 4: ...rehabilitation of numerous injuries The laser light penetrates tissue where it is absorbed by cells and effectively stimulates healing and relief of pain It also induces an immediate analgesic effect...

Page 5: ...D laser emission indicator light 5 Identification label 6 Battery cover 7 Danger label 8 Aperture warning label 9 Certification label 10 On Off switch 11 ENTER button 12 START STOP button 13 Reset but...

Page 6: ...ion indicator light 5 Identification Label 9 Certification Label 13 Reset buttons 2 Up button 3 Down button 12 Start Stop button 6 Battery cover 10 On Off switch button 11 Enter button 7 Danger Label...

Page 7: ...he name and the location of the Manufacturer Laserex Technologies Pty Ltd This label appears on the front of the unit Warning logotype The Danger label appears on the top of the unit and states the Ma...

Page 8: ...7...

Page 9: ...specified herein may result in hazardous radiation exposure Operating Instructions Charging Batteries Please refer to the Powercell leaflet included with the AEL IR 4400 1 Installation of batteries Re...

Page 10: ...k position by lining up the notch and turning it clockwise to the 12 o clock position 3 Operating the unit Press the Green On Off button for a few seconds An audible beep will be heard The following a...

Page 11: ...il the correct setting is highlighted and then press ENTER Continue Last From Data Source screen and with Continue Last highlighted in blue press ENTER and the most recent treatment protocol will appe...

Page 12: ...ed in orange Press UP to increase the dosage or DOWN to decrease When the correct dosage is entered press ENTER Frequency Selection Frequency Freq will be highlighted in blue and Continuous Wave CW wi...

Page 13: ...lowing screen appears As the AEL IR 4400 is a Class 3B laser it requires a RF safety interlock You will receive 2 of these with the unit These can be placed either on the door or window of your office...

Page 14: ...dy at the position where the subject is to receive treatment Press the red START STOP button and the AEL IR 4400 delivers the requested dosage An audible beep will be heard and the Laser Emission indi...

Page 15: ...memory In future when the stored dosage is required press ENTER at Load Memory at the Date Source screen If you arrive at the above screen and notice that an error was made press DOWN so that CANCEL...

Page 16: ...ettings press UP to select the setting highlighted in blue Then press ENTER to highlight the in orange Press UP to increase the value or DOWN to decrease Stopping the unit If at any time you wish to s...

Page 17: ...hed to the AEL IR 4400 the following will appear on the screen Press Enter with OK highlighted The next screen will indicate the following The AEL IR 4400 should be switched off and the head re attach...

Page 18: ...rongly recommended that laser safety glasses provided with the unit are worn during operation Laser Emission Indicator When the AEL IR 4400 is operating the laser emission indicator button will be fla...

Page 19: ...erse the device in water or liquids Any servicing of the device must be done through an Authorised Dealer If you have any queries refer them to Laserex Technologies WARNINGS Do Not use the AEL IR 4400...

Page 20: ...ilable 904nm x 3 635nm x 1 Power 450mW Pulse Repetition Mode Continuous Wave and preset Nogier frequencies Battery Type 4 x AA NiMH rechargeable batteries Battery Life Approximately 2 hours continuous...

Page 21: ...quality control Every effort has been made to ensure that the device reaches you in good working condition However if the device should malfunction due to defective materials and or workmanship it is...

Page 22: ...21 Address of Manufacturer Laserex Technologies Pty Ltd 136 Sturt Street Adelaide SA 5000 Australia Telephone 61 8 8221 6844 Facsimile 61 8 8221 6944 Web www laser3000 net E mail sales laserex net...

Page 23: ...22 Your Authorised Dealer is...

Reviews: