Chapter 3 — Measuring with the DWR-10
Page 44
September 2004
When sufficient readings have been recorded, you can
either stop the automatic measurements or test another
winding.
To stop recording measurements:
1. Press
either
the
End & Record Session
button on the
default toolbar,
or
the
Clock On/Off
button.
2. Press
the
Test/Dump
button.
To test another winding:
1. Press
the
Auto On/Off
button to stop recording (the
Next meas
. field will display
OFF
).
2. Press
the
Test/Dump
button.
3. Change the windings.
4. Press
the
Test/Dump
button again.
5. Edit
the
Tap
and
Winding
fields if time permits.
6. When readings are stable, press the
Auto On/Off
button to start recording.
Note: Any time during
Timed
mode, you can also press
the
Save
button to take additional manual measurements.
The
Save
button is available whether or not the Auto
feature is active.
All manuals and user guides at all-guides.com