Technical Specification
Page 24
Technical Specifications
Item
Description
Power Input
85 to 265VAC, 50/60Hz, 800VA max., CAT II
Fuse Rating
8A 250VAC, Type T
Test Current
From 10 to 100 A in 10A steps
Ripple <
0.5%
Test Current -
Permanent
20A
Test Time
From 10s to 60s in 1s steps (10-100A)
Infinite test time at 20A
Ramp time
2 to 3 seconds, related with selected current and
load
Resistance
Measurement
One auto-ranging channel
"Resistance
Ranges
0 µ
Ω
to 1999.9 µ
Ω
(10 to 100A)
0 m
Ω
to 19.999 m
Ω
(10 to 100A)
0 m
Ω
to 199.99 m
Ω
(10A)
Accuracy
±0.5% reading ±0.5% range
Resolution 4½
digits
Test Voltage
7V max.
Test Voltage
transients
26V max.
Test Current
transients
110A max.
Appendix