16
The temperature coefficient can be modified
through the “tc” setup item (see “Setup” section
for details).
• If the temperature reading is out of the -9.9 to
120.0°C range and the ATC option is selected,
the temperature full scale value and the °C tag
will blink on the secondary LCD.
• The reference temperature can be set from 15
to 30°C. When the reference temperature is
changed, the temperature coefficient must be
manually adjusted by the user.
For example, if
α
is the coefficient with refer-
ence temperature of 25°C, when changing the
temperature to 20°C, the new coefficient can
be calculated with the following formula:
β
=
α
/(1-
α
/20)
If
α
=1.90%/°C, then
β
=2.10%/°C.
• For AD332, always set reference tempera-
ture to 25°C when measuring TDS.
25
The following table lists the setup items, their valid range
and the factory settings (default):
Item
tc
tcE
rEF
tdS
CELL
AoFF
YEAr
DATE
TIME
id
vEr
Description
Temperature coefficient
Temp.compensation mode
Reference temperature
TDS factor (*)
Cell constant (K)
Auto-off feature
Printer Status (*)
Year
Date (DD.MM)
Time (hh:mm)
Meter identification code
Firmware release
Valid values Default
0.00 to 10.00% / ºC 1.90
Atc, Mtc, notc
Atc
15.0 to 30.0 ºC
25.0
ºC
0.40 to 1.00
0.50
0.500 to 1.700
1.000
On, Off
Off
On/Off
Off
2000 to 2098
2000
01.01 to 31.12
01.01
01:01 to 23:59
00:00
0000 to 9999
0000
(*) For AD332 only.