ADWA AD 310 User Manual Download Page 6

6

TECHNICAL DATA

EC Range

19.99 ; 199.9 ; 1999 

μ

S/cm

(AD310)

19.99 ; 199.9 mS/cm

TDS Range

9.99 ; 99.9 ; 999 ppm

(AD410)

9.99 ; 99.9 ppt

Temperature Range 

(both models)

-9.9 to 120.0 °C

Resolution

0.01 ; 0.1 ; 1 

μ

S/cm

0.01 ; 0.1 mS/cm

0.01 ; 0.1 ; 1 ppm / 0.01 ; 0.1 ppt

                                         0.1 °C

Accuracy

±1% f.s. (EC and TDS)

(@25 °C/77 °F)

±0.5 °C

EC Calibration

Offset: 0.0 

μ

S/cm;

(AD310)

Slope: 1 point with 6

memorized standards (84.0, 1413 

μ

S/cm; 5.00,

12.88, 80.0, 111.8 mS/cm) or one custom value

TDS Calibration

Offset: 0.0 ppm;

(AD410)

Slope: 1 point  with 4 memorized

standards (800, 1382 ppm; 6.44, 40.0 ppt)

or one custom value

23

Note:

Press SHIFT and then SETUP key before
confirmation to escape without changing the
previously set value.

The following table lists the setup items, their valid range
and the factory settings (default):

Description

Temperature coefficient
Temp. compensation mode
Reference temperature
Cell constant (K)
Auto-off enable
Firmware release

Valid values   

     

 Default

0.00 to 6.00% / ºC        1.90
Atc, Mtc, no

    Atc

20.0 or 25.0 ºC      25.0 

ºC

0.500 to 1.700             1.000
On, Off                       Off

Item

tc
tcE
rEF
CEL
AoF
vEr

Summary of Contents for AD 310

Page 1: ...EC TDS and Temperature w w w a d w a i n s t r u m e n t s c o m ADWA INSTRUMENTSKft Als kik t sor 11 6726 Szeged Hungary Tel 3662317878 Fax 3662550610 e mail sales adwainstruments com www adwainstru...

Page 2: ...mmences from the original date of sale Warranty is only valid when the product is used under normal conditions and in accordance with the instruction manual The warranty is void if the instrument is r...

Page 3: ...ent 25 Probe Maintenance 26 Probes and Solutions 27 TABLE OF CONTENTS 26 PROBE MAINTENANCE After measurements rinse the probe with clean water If a more thorough cleaning is required remove the probe...

Page 4: ...ser selectable The instruments also feature a measurement stability indicator 25 BATTERY REPLACEMENT When batteries become weak the battery symbol lights up It is recommended to replace the batteries...

Page 5: ...ation solution 20 ml sachet each 1413 S cm for AD310 1382 ppm for AD410 Batteries 4 x 1 5V AA alkaline Instruction manual 24 To freeze the reading on the LCD press the HOLD key from measurement mode T...

Page 6: ...ne custom value TDS Calibration Offset 0 0 ppm AD410 Slope 1 point with 4 memorized standards 800 1382 ppm 6 44 40 0 ppt or one custom value 23 Note Press SHIFT and then SETUP key before confirmation...

Page 7: ...vironment 0 to 50 C RH max 95 Dimensions 70 x 145 x 26 mm Weight Approx 200 g 22 SETUP Setup mode allows to view and modify the instrument parameters To enter setup mode press SHIFT and then SETUP key...

Page 8: ...laythefactorydefaulttemperature Adjust the temperature reading on the LCD using the arrow keys The maximum allowed adjustment is 1 0 C around current reading Press SHIFT and CFM keys to confirm The me...

Page 9: ...tivate the key alternate function press and hold first the SHIFT key and then the second desired key 8 Liquid Crystal Display LCD 20 When the reading is stable the hourglass tag stops blinking Press S...

Page 10: ...or a few seconds then the LCD displays the not calibrated temperature compensated EC or TDS reading Pressing SHIFT and RANGE keys the LCD will show the temperature compensated EC or TDS reading factor...

Page 11: ...theair Notes If temperature is out of range the buF message and the C tag blink on the LCD If the reading is too far from the expected value buF message blinks For best results calibrate using a stand...

Page 12: ...rature reading in the selected range Notes If the meter displays only dashes the reading is out of range Ifthestabilityindicator hourglasssymbol blinks the reading is not stable 17 If necessary use th...

Page 13: ...t samples for accurate reading it is recommended to rinse the probe thoroughly with deionized water before immersing it into the sample 16 CALIBRATION STANDARD CALIBRATION EC AD310 or TDS AD410 calibr...

Page 14: ...ompensation setting can be accessed by entering the setup mode and selecting the tcE item see Setup section for details 15 If the temperature compensation is active measurements are compensated using...

Reviews: