ADWA AD 310 User Manual Download Page 13

13

AUTORANGING

The EC (

AD310

) and TDS (

AD410

) scales are

autoranging. The meter automatically sets the scale with
the highest possible resolution.
By pressing SHIFT and FIXED keys,
the autoranging feature is disabled.
The current range is frozen on the
LCD and the tag corresponding to the
selected measurement range starts
blinking.
To restore the autoranging feature press the SHIFT and
FIXED keys again.

Note:

Autoranging is automatically restored if range is
changed, if setup or calibration mode is entered,
or if meter is turned off and back on again.

Make sure the meter is calibrated before taking
measurements.

If measurements are taken successively in
different samples, for accurate reading it is
recommended to rinse the probe thoroughly with
deionized water before immersing it into the
sample.

16

CALIBRATION

STANDARD CALIBRATION

EC (

AD310

) or TDS (

AD410

) calibration is a single

point procedure. Selectable points are 0.00, 84.0 and
1413 

μ

S/cm, 5.00, 12.88, 80.0 and 111.8 mS/cm for EC

calibration (

AD310

); 800, 1382 ppm and 6.44, 40.0 ppt

for TDS calibration (

AD410

).

• Rinse the probe with selected

calibration solution or deionized
water, then immerse it into the
solution. The sleeve holes must be
completely submerged.

• Tap the probe repeatedly to remove

any air bubbles that may be trapped
inside the sleeve.

• To enter calibration mode, press SHIFT

and then CAL from measurement
mode.

The LCD will show the “CAL” message for a few
seconds, followed by the not calibrated EC (

AD310

)

or TDS (

AD410

) reading together with blinking

hourglass symbol.

Summary of Contents for AD 310

Page 1: ...EC TDS and Temperature w w w a d w a i n s t r u m e n t s c o m ADWA INSTRUMENTSKft Als kik t sor 11 6726 Szeged Hungary Tel 3662317878 Fax 3662550610 e mail sales adwainstruments com www adwainstru...

Page 2: ...mmences from the original date of sale Warranty is only valid when the product is used under normal conditions and in accordance with the instruction manual The warranty is void if the instrument is r...

Page 3: ...ent 25 Probe Maintenance 26 Probes and Solutions 27 TABLE OF CONTENTS 26 PROBE MAINTENANCE After measurements rinse the probe with clean water If a more thorough cleaning is required remove the probe...

Page 4: ...ser selectable The instruments also feature a measurement stability indicator 25 BATTERY REPLACEMENT When batteries become weak the battery symbol lights up It is recommended to replace the batteries...

Page 5: ...ation solution 20 ml sachet each 1413 S cm for AD310 1382 ppm for AD410 Batteries 4 x 1 5V AA alkaline Instruction manual 24 To freeze the reading on the LCD press the HOLD key from measurement mode T...

Page 6: ...ne custom value TDS Calibration Offset 0 0 ppm AD410 Slope 1 point with 4 memorized standards 800 1382 ppm 6 44 40 0 ppt or one custom value 23 Note Press SHIFT and then SETUP key before confirmation...

Page 7: ...vironment 0 to 50 C RH max 95 Dimensions 70 x 145 x 26 mm Weight Approx 200 g 22 SETUP Setup mode allows to view and modify the instrument parameters To enter setup mode press SHIFT and then SETUP key...

Page 8: ...laythefactorydefaulttemperature Adjust the temperature reading on the LCD using the arrow keys The maximum allowed adjustment is 1 0 C around current reading Press SHIFT and CFM keys to confirm The me...

Page 9: ...tivate the key alternate function press and hold first the SHIFT key and then the second desired key 8 Liquid Crystal Display LCD 20 When the reading is stable the hourglass tag stops blinking Press S...

Page 10: ...or a few seconds then the LCD displays the not calibrated temperature compensated EC or TDS reading Pressing SHIFT and RANGE keys the LCD will show the temperature compensated EC or TDS reading factor...

Page 11: ...theair Notes If temperature is out of range the buF message and the C tag blink on the LCD If the reading is too far from the expected value buF message blinks For best results calibrate using a stand...

Page 12: ...rature reading in the selected range Notes If the meter displays only dashes the reading is out of range Ifthestabilityindicator hourglasssymbol blinks the reading is not stable 17 If necessary use th...

Page 13: ...t samples for accurate reading it is recommended to rinse the probe thoroughly with deionized water before immersing it into the sample 16 CALIBRATION STANDARD CALIBRATION EC AD310 or TDS AD410 calibr...

Page 14: ...ompensation setting can be accessed by entering the setup mode and selecting the tcE item see Setup section for details 15 If the temperature compensation is active measurements are compensated using...

Reviews: