TECHNICAL DATA
Range
0.00 to 19.99 µS/cm; 0.00 to 9.99 ppm
20.0 to 199.9 µS/cm;
10.0 to 99.9 ppm
200
to 1999 µS/cm; 100 to 999 ppm
2.00
to 19.99 mS/cm; 1.00 to 9.99 ppt
20.0 to 199.9 mS/cm; 10.0 to 99.9 ppt
-9.9 to 120.0°C
Resolution
0.01, 0.1, 1 µS/cm; ppm
0.01, 0.1 mS/cm; ppt
0.1°C
Accuracy
±1% f.s. (EC/TDS)
(@20°C/68°F)
±0.5°C
EC Calibration
Offset at 0.00 µS/cm;
Slope at 1 point
with 6 memorized values
(84.0, 1413 µS/cm; 5.00, 12.88, 80.0, 111.8 mS/cm)
or with custom value
Temperature
Automatic or manual,
Compensation
-9.9 to 120°C
Temperature
Selectable
Coefficient
from 0.00 to 10.00%/°C
TDS Factor
Selectable from 0.40 to 1.00
(default value: 0.50)
31
• time
• EC or TDS reading
(only dashes "----" means
reading out of range or
probe not connected)
• Temperature reading
(only dashes "----" means
reading out of range)
• To skip to the next or previous sample, use the up or
down arrow keys respectively.
For example, if the up arrow key is pressed while a
sample reading is displayed, the meter shows the read-
ing of next sample.
• To return to normal measurement mode, press the
SHIFT+RECALL keys at any time.