14
TEMPERATURE COMPENSATION
Three options are available for temperature compen-
sation:
1.
Automatic
(ATC): the EC probe features a built-in
temperature sensor, which provides the temperature
reading to automatically compensate the EC/TDS
measurement (from -9.9 to 120.0°C), also using the
selected reference temperature.
2.
Manual
(MTC): the temperature value can be manu-
ally set using the arrow keys. The compensation is
referenced to the selected reference temperature.
While in MTC mode, the °C tag blinks on the second-
ary LCD.
3.
No compensation
(NOTC): the temperature is not
taken into account. The reading displayed on the pri-
mary LCD is the actual EC or TDS value.
Notes:
• The default compensation mode is ATC.
• Temperature compensation setting can be ac-
cessed by entering the setup mode and selecting
the "tcE" item (see “Setup” section for details).
23
Press the arrow keys to change the value.
If there is another part of the item to be set (e.g. month
in setting up the current date), press RANGE and the
part to be changed will start blinking.
Press the ARROW keys to change the value and CFM
to confirm.
Note:
Press SETUP key before confirmation to escape
without changing the previously set value.