7.6 Marker Function
7-75
Target menu
{TARGET VALUE}
: Searches the specified value (response value).
If SMITH or POLAR is set in the format menu, LOG MAG
type of data is searched as TARGET VALUE.
However, if SMOOTHING is set to ON, the active
marker does not move to a correct data.
{ 0° }
: Searches the phase 0
°
.
The phase data is surely searched without regard to any
format.
If SMOOTHING is set to ON, the active marker does not
move to a correct data.
{
±
180°}
: Searches the phase 180
°
.
The phase data is surely searched without regard to any
format.
If SMOOTHING is set to ON, the active marker does not
move to a correct data.
{LEFT SEARCH}
: Searches specified value of left side from current marker
position.
{RIGHT SEARCH}
: Searches specified value of right side from current
marker position.
CAUTION!
1. When the internal data is searched, the searched position may not be the same as the
display because the smoothing is performed to the internal data when SMOOTHING is
ON.
2. TARGET VALUE can be specified only with the ten-key.
Ripple menu
{MAX }
: Searches for the maximum of local maximum peak
values.
When FORMAT is SMITH/POLAR, the internal
LOGMAG data is searched.
{MIN }
: Searches for the minimum of local minimum peak
values.
When FORMAT is SMITH/POLAR, the internal
LOGMAG data is searched.
4
5