R3267 Series OPT11 3GPP Level Calibration Operation Manual

5.2 Performance Verification Test Procedure

5-4

5.2

Performance Verification Test Procedure

5.2.1

Level Calibration

1.

Description

1.

Measure level calibration measurement accuracy for 3GPP system.

2.

Firstly measure a signal generator output signal level for several testing points by
using RF power meter and RF power sensor as reference data.

3.

Secondary, measure the signal generator output signal by Tx power measurement
function of R3267 Series.

4.

Compare the both measurement result of RF power meter and R3267 Series.

5.

Record the difference as level calibration measurement accuracy.

6.

Repeat measurement for input attenuator setting 10 dB through 50 dB. 
R3267 Series is phase locked with the signal generator.

2.

Specification

At 25 

°

C, After gain calibration performed, ATT=AUTO, MinATT=ON

Measurement Accuracy:

±

0.4 dB

(Input signal level: +25 dBm to -50 dBm)

3.

Equipment used

Signal Generator

:SG

RF Power Meter

:PM

RF Power Sensor

:PS

RF Cable

:CBL_1

RF Cable

:CBL_2

RF Cable

:CBL_3

Adapter

:Adap_1

Adapter

:Adap_2

Adapter

:Adap_3

Summary of Contents for R3264

Page 1: ...MBER C Printed in Japan ADVANTEST CORPORATION All rights reserved R3267 Series OPT11 3GPP Level Calibration Operation Manual FOE 8370635B01 First printing December 25 2000 Applicable R3264 R3267 R3273 2000 Cover ...

Page 2: ...otentially hazardous situation which will result in personal injury or a damage to property including the product Basic Precautions Please observe the following precautions to prevent fire burn electric shock and personal inju ry Use a power cable rated for the voltage in question Be sure however to use a power cable conforming to safety standards of your nation when using a product overseas When ...

Page 3: ...roduct or instrument or relat ing to a restriction on operation Safety Marks on the Product The following safety marks can be found on Advantest products ATTENTION Refer to manual Protective ground earth terminal DANGER High voltage CAUTION Risk of electric shock Replacing Parts with Limited Life The following parts used in the instrument are main parts with limited life Replace the parts listed b...

Page 4: ...back ups of important data The data stored in the disk may become damaged if the product is mishandled The hard disc has a limited life span which depends on the operational conditions Note that there is no guarantee for any loss of data Precautions when Disposing of this Instrument When disposing of harmful substances be sure dispose of them properly with abiding by the state provided law Harmful...

Page 5: ...ure 3 Storage Position This instrument can be used safely under the following conditions Altitude of up to 2000 m Installation Categories II Pollution Degree 2 Vibration Direct sunlight Corrosive gas Dust Front Keep at least 10 centimeters of space between the rear panel and any other surface Front This instrument should be stored in a horizontal position When placed in a vertical upright position...

Page 6: ...3 1 3 2 Menu Map 3 2 3 3 Functional Description 3 2 3 3 1 STD Menu 3 2 4 REMOTE PROGRAMMING 4 1 4 1 GPIB Command Index 4 1 4 2 GPIB Command Codes 4 2 5 PERFORMANCE VERIFICATION TEST 5 1 5 1 GENERAL 5 1 5 1 1 Introduction 5 1 5 1 2 Test Equipment 5 1 5 1 3 Performance Verification Test Record Sheet 5 2 5 1 4 Performance Verification Procedure 5 3 5 2 Performance Verification Test Procedure 5 4 5 2 ...

Page 7: ...el Calibration Operation Manual LIST OF ILLUSTRATIONS No Title Page 1 1 Installed Option Labels 1 1 2 1 2 GHz Modulation Signal Measurement Sample Using the 3GPP Level Calibration Function 2 4 5 1 Setup of Level Calibration Test 5 5 ...

Page 8: ...Page 2 1 Frequency Range for 3GPP Level Calibration Data and the Frequency 2 1 2 2 Setting Conditions Used for 3GPP Level Calibrations 2 1 4 1 3GPP Level Calibration OPT11 4 2 5 1 Equipment List 5 2 5 2 List for SG and PM Correction Data Setting 5 6 5 3 Setting of SG and R3267 Series 5 7 ...

Page 9: ... in the 2 GHz band 1 8483 GHz to 2 1717 GHz when the 3GPP level calibration option OPT 11 is installed in the R3267 Series Spec trum Analyzer 1 2 Accessories 1 3 Precautions in Use 1 Option labels The number 11 is written on the INSTALLED OPTION NO label attached to the rear panel when the 3GPP calibration option is installed in the R3267 Series Figure 1 1 Installed Option Labels Name of accesorie...

Page 10: ... procedure below before making measurements 1 Connect the A01036 0150 cable that comes with the R3267 Series to the 30 MHz CAL OUT connector and the INPUT connector 2 Press SHIFT 7 CAL and Cal All Cal All is performed this takes approx 9 minutes b Performing Gain Cal Performs a calibration for the Tx power measurement Perform this calibration when an ambient temperature changes 1 Connect the A0103...

Page 11: ...3GPP level calibration data and the frequency interval between cal ibrations Table 2 2 shows the setting conditions used for 3GPP level calibrations Table 2 1 Frequency Range for 3GPP Level Calibration Data and the Frequency Frequency range Frequency interval between calibrations 1 8483 GHz to 2 1717 GHz 6 6 MHz Table 2 2 Setting Conditions Used for 3GPP Level Calibrations Item Setting Input atten...

Page 12: ...hether or not 3GPP level calibration data is used NOTE The initial setting after the PRESET sets Level Cal Correction to ON 1 Press TRANSIENT and STD The menu that includes STD is displayed 2 Press Level Cal Correction ON OFF ON 3GPP level calibration data is used for the measurement OFF 3GPP level calibration data is not used for the measurement ...

Page 13: ...ition 2 Press FREQ 2 and GHz The center frequency is set to 2 GHz Level calibration for the R3267 Series 3 Connect the A01036 0150 cable that comes with the R3267 Series to the 30 MHz CAL OUT connector and the INPUT connector 4 Press TRANSIENT STD and Gain Cal Gain Cal is performed this takes approx 20 seconds The level is automatically calibrated for the R3267 Series 5 Input a modulation signal t...

Page 14: ...ual 2 3 Measurement Sample 2 4 7 Press SINGLE The modulation signal power is measured The level of the modulation signal is displayed on the R3267 Series screen Figure 2 1 2 GHz Modulation Signal Measurement Sample Using the 3GPP Level Calibration Function ...

Page 15: ...s 3 REFERENCE This chapter describes the keys used with the 3GPP level calibration option OPT 11 For information on the other keys refer to Section 3 3 Functional Description in R3267 Series Spectrum Analyzer Operation Manual 3 1 Menu Index Level Cal Correction ON OFF 3 2 STD 3 2 TRANSIENT 3 2 ...

Page 16: ...vel calibration option OPT 11 3 3 1 STD Menu Press the TRANSIENT and STD keys displays Level Cal Correction ON OFF is displayed Level Cal Correction ON OFF Toggles the level correction function on or off ON 3GPP level calibration data is used for the measure ment OFF 3GPP level calibration data is not used for the measure ment TRANSIENT STD Level Cal Correction ON OFF ...

Page 17: ...4 REMOTE PROGRAMMING 4 1 GPIB Command Pages 4 REMOTE PROGRAMMING A list of GPIB codes required for remote control is shown in this section 4 1 GPIB Command Index This GPIB command index can be used as the index for this Chapter LCALCORR OFF 4 2 LCALCORR ON 4 2 ...

Page 18: ...Codes 4 2 4 2 GPIB Command Codes The following table list the GPIB commands by function Table 4 1 3GPP Level Calibration OPT11 Function Listener Code Talker Request Code Output Format Level Cal Correction Level Cal Correction ON LCALCORR ON LCALCORR 0 OFF OFF LCALCORR OFF 1 ON ...

Page 19: ... Test Equipment The Table 5 1 lists recommended test equipment The equipment needed to perform all of the performance test Equipment lists for individual tests are provided in each performance verification test CAUTION 1 The R3267 Series with OPT 11 to be tested should be warm up for at least 60 minutes before starting test 2 Make sure that the test equipment used meets its own published specifica...

Page 20: ...odel Manufacturer Notes 1 RF Power Meter RF Power Sensor Frequency Range 10 MHz to 3 0 GHz Power Range 1 µW to 100 mW VSWR 1 1 Measurement accuracy 0 08 dB Note Must be calibrated in advance NRVS NRV Z52 Rohde Schwarz PM PS 2 Signal Genera tor Frequency Range 10 MHz to 3 0 GHz Output Level 15 dBm to 10 dBm Aging Rate 1 x 10 exp 6 year SMP03 equipped Option B11 Rohde Schwarz SG 3 RF Cable Impedance...

Page 21: ... the text as follows Panel keys Boldface type Example FREQ FORMAT Soft keys Boldface and ItalicExample Center Trace Detector When a series of key operations is described using a comma between two keys There are various soft menus used to switch between two states such as ON OFF and AUTO MNL For example when turning off the Display ON OFF function the annotation Display ON OFF OFF is used When swit...

Page 22: ...rator output signal by Tx power measurement function of R3267 Series 4 Compare the both measurement result of RF power meter and R3267 Series 5 Record the difference as level calibration measurement accuracy 6 Repeat measurement for input attenuator setting 10 dB through 50 dB R3267 Series is phase locked with the signal generator 2 Specification At 25 C After gain calibration performed ATT AUTO M...

Page 23: ... Frequency 1 8483 GHz 5 On the PM set a correction data for 1 8483 GHz 6 While monitoring PM measurement data adjust SG output level so that measure ment data to be 12 0 1 dBm 7 Record the PM measurement data in the PM column on the performance verifi cation test record sheet for attenuator setting 10 dB through 50 dB 8 On the SG set control as follow Output Frequency 1 9011 GHz 9 Record the PM me...

Page 24: ...s INPUT by using CBL_1 and Adap_3 as shown dashed line in Figure 1 INPUT ATT 10 dB Setting 16 On the R3267 Series press keys as follows to set an input attenuator to 10 dB ATT ATT AUTO MNL MNL 1 0 GHz dB 17 On the SG set control as follow Output Frequency 1 8483 GHz 18 On the R3267 Series set control as follow Center Frequency 1 8483 GHz 19 On the R3267 Series press keys as follows to perform auto...

Page 25: ...2 0 GHz dB 25 Repeat steps 17 through 23 INPUT ATT 30 dB Setting 26 On the R3267 Series press keys as follows to set the input attenuator to 30 dB ATT ATT AUTO MNL MNL 3 0 GHz dB 27 Repeat steps 17 through 23 INPUT ATT 40 dB Setting 28 On the R3267 Series press keys as follows to set the input attenuator to 40 dB ATT ATT AUTO MNL MNL 4 0 GHz dB 29 Repeat steps 17 through 23 INPUT ATT 50 dB Setting...

Page 26: ...ult Frequency GHz Min dB Measured error dB Max dB Pass Fail 1 8483 0 4 0 4 1 9011 0 4 0 4 1 9473 0 4 0 4 2 0001 0 4 0 4 2 0529 0 4 0 4 2 0991 0 4 0 4 2 1519 0 4 0 4 2 1717 0 4 0 4 ATT 30 dB Measured value of power meter dBm R3267 Series Tx Power Measured value dBm Specification Result Frequency GHz Min dB Measured error dB Max dB Pass Fail 1 8483 0 4 0 4 1 9011 0 4 0 4 1 9473 0 4 0 4 2 0001 0 4 0 ...

Page 27: ...cord Sheet 5 9 ATT 50 dB Measured value of power meter dBm R3267 Series Tx Power Measured value dBm Specification Result Frequency GHz Min dB Measured error dB Max dB Pass Fail 1 8483 0 4 0 4 1 9011 0 4 0 4 1 9473 0 4 0 4 2 0001 0 4 0 4 2 0529 0 4 0 4 2 0991 0 4 0 4 2 1519 0 4 0 4 2 1717 0 4 0 4 ...

Page 28: ...bration frequency range 1848 3 MHz to 2171 7 MHz Level measurement range 25 dBm to 60 dBm Level measurement accuracy Measurement error When ATT is set to AUTO and Min ATT is set to ON after Gain Cal is per formed at 25 C Measurement linearity 0 4 dB 25 dBm to 50 dBm 0 6 dB 50 dBm to 60 dBm 0 2 dB 0 dB to 30 dB Gain Cal error due to temperature 0 015 dB C Calibration period 1 year ...

Page 29: ...R3267 Series OPT11 3GPP Level Calibration Operation Manual A 1 APPENDIX A 1 Error Message List Code Error message Description 752 Invalid Level Cal Correction data Contact qualified engineer ...

Page 30: ...1 I INTRODUCTION 1 1 L Level Cal Correction ON OFF 3 2 Level Calibration 5 4 M Measurement Sample 2 3 Menu Index 3 1 Menu Map 3 2 O OPERATION 2 1 P Performance Verification Procedure 5 3 PERFORMANCE VERIFICATION TEST 5 1 Performance Verification Test Procedure 5 4 Performance Verification Test Record Sheet 5 2 5 8 Precautions in Use 1 1 Product Description 1 1 R REFERENCE 3 1 REMOTE PROGRAMMING 4 ...

Page 31: ... hereby grants you a license to use the SOFTWARE only on or with ADVANTEST PRODUCTS Restrictions 1 You may not use the SOFTWARE for any purpose other than for the use of the PRODUCTS 2 You may not copy modify or change all or any part of the SOFTWARE without permission from ADVANTEST 3 You may not reverse engineer de compile or disassemble all or any part of the SOFTWARE Liability ADVANTEST shall ...

Page 32: ...mitation i instances where the Product has been subjected to physical stress or electrical voltage exceeding the permissible range and ii instances where the corrosion of electrical circuits or other deterioration was accelerated by exposure to corrosive gases or dusty environments d use of the Product in connection with software interfaces products or parts other than software interfaces products...

Page 33: ...s expired and such component or part has not been replaced there is a possibility that the Product will not perform properly Additionally if the operating life of a component or part has expired and continued use of such component or part damages the Product the Product may not be repairable Please contact the nearest ADVANTEST office listed at the end of this Operation Manual or ADVANTEST s sales...

Page 34: ... 49 89 4129 13723 Advantest Singapore Pte Ltd Singapore 438A Alexandra Road 8 03 06 Alexandra Technopark Singapore 119967 Phone 65 6 274 3100 Facsimile 65 6 274 4055 Advantest Korea Co Ltd Korea 16Fl MIRAEWASARAM Bldg 942 1 Daechi Dong Kangnam Ku Seoul Korea Phone 82 02 3452 7710 Facsimile 82 02 3452 7970 Advantest Suzhou Co Ltd China 5F No 46 Factory Building No 555 Gui Ping Road Shanghai China 2...

Reviews: