The specifications for the 3GPP level calibration option (OPT 11) are shown below.
Calibration frequency range
1848.3 MHz to 2171.7 MHz
Level measurement range
+25 dBm to -60 dBm
Level measurement
accuracy
Measurement error
(When ATT is set to AUTO and Min
ATT is set to ON after Gain Cal is per-
formed at 25
°
C)
Measurement linearity
≤ ±
0.4 dB (+25 dBm to -50 dBm)
≤ ±
0.6 dB (-50 dBm to -60 dBm)
≤ ±
0.2 dB (0 dB to -30 dB)
Gain Cal error due to temperature
0.015 dB/
°
C
Calibration period