R3131 Spectrum Analyzer Operation Manual
2.4 Other Functions
2-89
Aug 28/98
Selecting the data to be saved
3.
Press Save Item.
The Save Item menu is displayed.
4.
Select the data you want to save from the Save Item menu.
Setup ON/OFF
: current settings
Trace ON/OFF
:
501-point trace A (or B), or data for both trace A and B
Ant Corr ON/OFF :
correction data
Norm Corr ON/OFF: Normalization calibration data (available only when
equipped with Option 74).
Trc Lvl ON/OFF
:
level values for the trace data (available when trace data
is being saved)
Figure 2-85 Display for Selecting the Data to be Saved
5.
Press RETURN.
Returns to the Save menu.
Choosing the file name
6.
Select the file name you want to save your data under using the data knob or step
keys. When the file is being saved to RAM, the file names appear as REG01 or
above. For floppy disks, the file names start from FILE00 and continue with 01,
02 etc.