21
iPS-M S series User Manual V.A1
Chapter 4
E
MC Table
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user
of the model IPS-M Bat-
tery requires continued
operation during power
mains interruptions, it is
recommended that the
model IPS-M Battery be
powered from an uninter-
ruptible power supply or
another battery.
NOTE
U
T is the a.c. mains voltage prior to application of the test level.