![Advanced Instruments 4250 User Manual Download Page 75](http://html2.mh-extra.com/html/advanced-instruments/4250/4250_user-manual_3047840075.webp)
Replacement,
”
shown later in this chapter.
Record the sample probe resistance and bin number in the service log at
the end of this user’s guide (and/or other convenient location); then press
the
STOP
key. The display will revert to
“
Probe Bin Test
”
.
Press
SETUP
and use the procedure
“
Set Sample Bin #
”
described in
Chapter 2 to reset the sample probe bin setting as necessary.
NOTE
Recalibration is required each time the sample probe is replaced
and/or the sample probe bin number is reset.
If running this test produces the message,
“
Reset Probe Configuration
”
instead of the sample probe resistance and bin number, the sample probe
bin number is not available from the instrument memory and must be
reset. Press
SETUP
and use the procedure
“
Set Sample Bin #
”
described
in Chapter 2 to preset a random sample probe bin number, such as
“
5
”
.
Then press
TEST
and rerun the probe bin test. Finally, as necessary,
press
SETUP
and use the
“
Set Sample Bin #
”
procedure to correct the
sample probe bin setting. If the
“
Reset Probe Configuration
”
message
persists, please refer to Chapter 5
“
Obtaining Service
”
to get help.
6. Barcode Test
This test performs a continuous check of the barcode port.
Press the
START
key to enter the test. At
“
[START] Test [STOP]
”
you
may press
START
once the barcode scanner is connected, or press
STOP
to return to the main menu.
After you press
START
you may begin scanning barcodes. If your code
will not scan, check the manufacturers user’s guide to verify correct
setup of the scanner for your symbologies.
Press
STOP
to exit to the test menu.
NOTE
This test will only work after power-on diagnostics have been
completed.
53
Troubleshooting & Service
Summary of Contents for 4250
Page 1: ...4255 Rev11 032317 The Advanced Cryoscope Model 4250 User s Guide...
Page 6: ...Notes vi Advanced Cryoscope Model 4250 User s Guide...
Page 14: ...xiv Advanced Cryoscope Model 4250 User s Guide Notes...
Page 28: ...Advanced Cryoscope Model 4250 User s Guide 6 Figure 3 Load Printer Paper...
Page 56: ...34 Advanced Cryoscope Model 4250 User s Guide Notes...
Page 60: ...Advanced Cryoscope Model 4250 User s Guide 38 Notes...
Page 88: ...66 Advanced Cryoscope Model 4250 User s Guide Notes...
Page 102: ...80 Advanced Cryoscope Model 4250 User s Guide Notes...
Page 106: ...Advanced Cryoscope Model 4250 User s Guide 84 Notes...
Page 110: ...88 Advanced Cryoscope Model 4250 User s Guide Notes...
Page 114: ...92 Advanced Cryoscope Model 4250 User s Guide Notes...
Page 118: ...Advanced Cryoscope Model 4250 User s Guide 96 Date Problem Symptom Action...