41
Troubleshooting Table
“
Set Probe Bin #'s
”
“
Sample Did Not
Freeze...
”
“
Solenoid Driver
Failure
”
“
Standards Reversed?
Please Repeat...
”
“
System Error: Reqs
”
“
System Error: Prim
”
“
System Error: Comm
”
“
System Error: Unkn
”
“
System Error: Trap
”
Your instrument will not function unless
there are bin numbers entered for both the
block and sample probes. To set these num-
bers, see instructions in Chapter 4.
If you heard an impact while the sample
was attempting to run, the fault most likely
lies with the sample preparation or the
block probe. Run another test, paying close
attention to technique. If your problem is
not resolved, check and correct if necessary
the block and probe bin numbers.If you
heard no impact, the problem most likely
lies with the solenoid impactor. Retry the
test, and, if necessary, contact Advanced
Instruments for Hot-Line Service.
This message indicates a problem with your
solenoid impactor. Try restarting your
instrument. If this does not solve your prob-
lem, contact Advanced Instruments for Hot-
Line Service.
This message will appear during the cali-
bration procedure if the instrument detects
that the low and high calibration standards
may have been mixed up and were entered
in the wrong order. Retry the calibration,
being sure that the standards are correct.
Each of these error messages indicate a differ-
ent system error. Try restarting your instru-
ment. If this does not solve your problem, con-
tact Advanced Instruments for Hot-Line
Service.
Summary of Contents for 3300
Page 1: ...The Advanced Micro Osmometer Model 3300 User s Guide 3305 Rev9 012303 ...
Page 6: ...vi Notes ...
Page 8: ...viii Notes ...
Page 12: ...xii Advanced Osmometer Model 3300 User s Guide Notes ...
Page 18: ...xviii Advanced Osmometer Model 3300 User s Guide Notes ...
Page 42: ...Advanced Osmometer Model 3300 User s Guide 24 Notes ...
Page 54: ...36 Advanced Osmometer Model 3300 User s Guide Notes ...
Page 68: ...Advanced Osmometer Model 3300 User s Guide 50 Notes ...
Page 72: ...Advanced Osmometer Model 3300 User s Guide 54 Date Problem Symptom Action ...
Page 75: ... Chapter Name ...
Page 76: ... Advanced Osmometer Model 3300 User s Guide ...
Page 77: ... Chapter Name ...