6-2
Troubleshooting and Global Customer Support
5705211-C
Advanced Energy
®
4. Is the
Interlock OK
indicator on?
a. If
yes
b. If
no
, see
“Interlock OK Indicator Off” on page 6-5
.
5. Is the
Load Regulation
indicator on solid or flashing?
a. If
solid
, go to step 6.
b. If
flashing
, see
“Load Regulation Indicator Flashing” on page 6-5
6. Is the
RF On
indicator lit?
a. If
yes
b. If
no
, see
“RF On Indicator Off” on page 6-6
.
7. Is the
At Set Point
indicator on?
a. If
yes
, it indicates that the HFV generator is operating properly and delivering
the requested power level.
b. If
no
, see
“At Set Point Indicator Off” on page 6-7
If these questions do not help solve your problem, contact AE Global Customer
Support (see
“AE Global Customer Support” on page 6-10
).
ERROR INDICATIONS
Front Panel Display (LCD) Not Lit
The front panel display should operate as soon as the rear circuit breaker is turned on.
Verify that the:
1. Rear circuit breaker is on.
2. Input line voltage is between 178 to 229 VAC between all three phases.
Advanced Energy Displayed on LCD
It is normal for ADVANCED ENERGY to display for a few seconds immediately
after the rear circuit breaker has been enabled (this is not an internal fault mode
indicator). However, if the message continually displays, try cycling the unit's circuit
breaker. If the problem persists, contact AE Global Customer Support (see
Summary of Contents for HFV 8000
Page 1: ...HFV 8000 Generator 2 MHz 208 V Input with DeviceNet March 2004 5705211 C User Manual...
Page 2: ......
Page 3: ...User Manual HFV 8000 Generator 2 MHz 208 V Input with DeviceNet 5705211 C...
Page 6: ...iv 5705211 C Advanced Energy...
Page 14: ...xii 5705211 C Advanced Energy...
Page 22: ...1 8 Safety and Product Compliance Guidelines 5705211 C Advanced Energy...
Page 26: ...2 4 Product Overview and Theory 5705211 C Advanced Energy...
Page 36: ...3 10 Product Specifications 5705211 C Advanced Energy...
Page 94: ...4 58 Communication Interfaces 5705211 C Advanced Energy...
Page 114: ...5 20 Installation Setup and Operation 5705211 C Advanced Energy...
Page 128: ...6 14 Troubleshooting and Global Customer Support 5705211 C Advanced Energy...
Page 132: ...iv 5705211 C Advanced Energy...
Page 134: ...5705211 C ii HFV 8000 Generator...
Page 140: ...5705211 C viii HFV 8000 Generator...
Page 141: ...ix 5705211 C Advanced Energy...
Page 142: ...Index x 5705211 C Advanced Energy...