ADOS S.R.L. Buccinasco (MI)
492-MTU10006
Rev. 4
Page 45 of 49
6.4.1
Zero calibration
The weighing system must be set with ZERO weight.
After selecting the modality:
C A L
O
press the
E
key.
The instrument prepares to receive the confirmation command and flashes the message on the display.
Press the
E
key again and the instrument starts the zero calibration procedure. The display shows the
waiting time:
W A I T
5
During the waiting time the load cell signal is integrated and it must stay stable.
When the wait time expires instrument performs the zero calibration and displays the following message for
2 seconds:
D O N E
and returns automatically to the calibration menu.