eP2 Controller

 

Operators Manual

 

4.1 The Test Method Menu 

1)

 

Set Test Method - 

Toggle the 1) key to select one of the six test methods. The active method is displayed on the Live Indicating 

Screen in the upper left hand corner of the LCD. For example, "M2", would mean that test method 2 is active. A test method includes 
all of the user-specified parameters in the Data Acq, Analysis and Control menus. In general, it is used to define how a test is run and 
what results are calculated and displayed.  

2)

 

Set ABS Zero Pos- 

Usually used in compression test to define absolute zero position as the point when the two platens touch. This 

is required for 

free height

 measurements. The absolute zero position is stored to permanent memory so that it is remembered on 

power cycles. The plates of the testing machine move till they reach a preset force value, then retract to the user set Home position. 
The Absolute Zero Function will determine the zero position to be used for subsequent measurements. 

4.2 The Data Acq Menu 

Press the <Data Acq> key with the LCD in the Live Indicating mode to enter the Data Acq Menu.  Use the <ARROW> keys to scroll to 

other menu items.  Use the number keys to select options. The following functions are provided in the Data Acq Menu. 

1)

 

Threshold

 - Set the Start Test Threshold in units of force.  Data logging for test starts when load crosses threshold. Once the 

START Button is pressed on the front panel and the load exceeds the Start Test Threshold the Test Status will change to "T" and 
the eP2 Controller will begin logging data. 

2)

 

Sample Break

 - Used to specify the end of test. User specifies sample break as a percentage drop from maximum load. The 

controller must record the percentage drop on three consecutive collection points before a break is registered. Setting sample 
break to 100% disables it. 

3)

 

Specimen Type

 - Set the specimen geometry and dimensions for stress calculations. Select specimen geometries by tapping the 

#1 key, choices are NONE, ROUND, FLAT,  BEAM CTR, BEAM CTR - RECT TUBE, BEAM 3RD,  AREA and WIDTH. After selecting the 
shape of the specimen you are given the opportunity to enter the specimen's dimensions. Dimension selections vary with 
specimen shape (type). 

4)

 

Log Rate - 

Set the data logging rate/test length AND specify whether to stream XY data to USB during test. Change log rate/test 

length by pressing 1) key: 

Logging Rate 

Maximum Logging Duration 

1000 Hz (1000/second) 

15 seconds 

500 Hz (500/second) 

30 seconds 

250 Hz (250/second) 

60 seconds 

100 Hz (100/second) 

150 seconds 

50 Hz (50/second) 

5 minutes 

25 Hz (25/second) 

10 minutes 

10 Hz (10/second) 

25 minutes 

5 Hz (5/second) 

50 minutes 

2 Hz (2/second) 

125 minutes 

1 Hz (1/second) 

250 minutes 

1

2

 Hz (30/minute) 

500 minutes 

1

6

 Hz (10/minute) 

25 hours 

1

30

 Hz (2/minute) 

125 hours 

 Toggle the 2) key to turn live streaming of XY data ON/Off.  

Summary of Contents for eP2

Page 1: ...51 Morgan Drive Norwood MA 02062 1 Tel 781 769 0850 Fax 781 769 0884 Sales ADMET com www ADMET com R7 10 2018 eP2 Controller Operators Manual...

Page 2: ...ided the buyer gives The Company prompt notice of any defect or failure and satisfactory proof thereof Products may be returned to The Company s factory The Company will deliver replacements for defec...

Page 3: ...written materials accompanying the eP2 CONTROLLER 3 0 Restrictions This Agreement and the Licensed Software Firmware may not be assigned Sublicensed or otherwise transferred by the Customer The end us...

Page 4: ...Key 9 3 7 E STOP RED BUTTON ON TOP OF UNIT 9 4 0 THE MENUS and KEYS 10 4 0 1 Test Method Specific Menu Selections 10 4 0 2 Global Menu Selections 11 4 1 The Test Method Menu 12 4 2 The Data Acq Menu 1...

Page 5: ...t you have become an owner We welcome your comments about our products and wish that you express them It is the only way that we can continue to build the best available test systems to satisfy your n...

Page 6: ...er travel protection is provided by means of independent limit switches Dual Segment control provides a means of more complex control including a true hold force feature and Cyclic control provides th...

Page 7: ...want the machine to move during test 2 Run Test Move crosshead to appropriate position using Jog keys and then set Home position and or zero load by using the Zero menu Press START button to begin th...

Page 8: ...ide of the Live Indicating display there are two letter codes Calibration Table The active load calibration table is displayed in the upper left of the live screen The eP2 digital controller holds up...

Page 9: ...with the data Note that this takes little time for most tests and may never actually be seen by the operator D Done Data collection has stopped X Buffer Full Data logged during tests exceeds the memor...

Page 10: ...CURRENT TEST pressing ENTER will cause the current test results to be printed Turns ON OFF a menu option Starts new test when controller in ready status 3 4 The ESC STOP Key Returns the LCD to the Li...

Page 11: ...Threshold 2 Sample Break 3 Specimen Type 4 Log Rate 5 Local Overload 6 Engineering Units Control Up Down Arrows to scroll through menu 1 Test Profile SegA in min End CH SegA End AmtA Create SegB Yes...

Page 12: ...ure specific UTILS Up Down Arrows to scroll through menu 1 Clear Results 2 Store Auto Manual 3 Set Date Time 4 Stiffness Factor 5 Install Defaults 6 Diagnostics 7 Transducer 8 Supervisor On Off PRINT...

Page 13: ...nu 1 Threshold Set the Start Test Threshold in units of force Data logging for test starts when load crosses threshold Once the START Button is pressed on the front panel and the load exceeds the Star...

Page 14: ...ENT key will return the focus to the main Analysis menu If any of the active analyses require setup parameters then the program will prompt for them accordingly For example if the Average Load analysi...

Page 15: ...t specified break point Load Change None Percent Elong Percent Elongation length at specified break point divided by original gage length Load Change Gage Length Load at Disp Load at specified displac...

Page 16: ...mber keys to select options The following functions are provided in the UTILS Menu 1 Clear Results Clear all stored results for the active test method or for all test methods 2 Store Set to store test...

Page 17: ...for segment B and press ENT The rate will be in units appropriate to control channel For example if CONTROL CH SegB is load then the rate units will be force time and if the CONTROL CH SegB is Positi...

Page 18: ...he home location is relative to zero position NOTE The absolute position zero is not affected by this action 4 Strain Resets the strain readings to zero only appears if strain option is installed 5 0...

Page 19: ...ly active engineering units Step 9 Input the resolution minimum load increment Step 10 Input the transducer serial number Step 11 Calibrate to enter the measure mode or ESC to exit NOTE If you decide...

Page 20: ...nding engineering unit value for each measurement point After all values have been entered the program will automatically generate the corresponding scale factors and store them to permanent memory DO...

Page 21: ...le Break are set to values appropriate for your test Threshold is where data logging begins Once the threshold is crossed the software is looking for the load to drop fromthe maximum load as specified...

Page 22: ...tuning gains for position control have been set at the factory The tuning gains for load control need to be set by the customer as these are dependent on the sample being tested Incorrect Test Method...

Page 23: ...Dimensions 8 0 Rear Panel Connections Optional Rear of eP2 7600 5600 R Rear of eP2 2600 Figure 8 0 1 Rear Panel Connections 9 25 4 00 8 50 Emergency Stop Actuator Connection Power from Frame USB Comp...

Page 24: ...put optional Figure 8 0 2 Analog Transducer Pin Designation Encoder Excitation pins 7 8 Encoder Input pins 14 15 Encoder shield connects to case Transducer Identification Contact Factory Servo Box Con...

Page 25: ...0850 Fax 781 769 0884 Sales ADMET com www ADMET com R7 10 2018 eP2 Controller Operators Manual 9 0 Factory Passwords Factory Passwords 268 calibration 157 supervisor mode Suggestion Remove this page...

Reviews: