15
7.4 Device configuration commands
Configuration commands are used to set parameters of the Atlas that are
used by the measurement functions. The settings are used globally by
measurement functions.
Command
Parameters
Purpose
:SENSe:INT
1: device
2: int time
Select device, 2D, spectrometer
or luminance meter and set
integration time
:SENSe:AVERag
e
1: device
2: average
Select device, 2D, spectrometer
or luminance meter and set
average
:SENSe:AUTOra
nge
1: device
2: on/off
Select device, 2D, spectrometer
or luminance meter and set auto-
range true or false
:SENSe:MAXInt
1: device
2: Max Int time
Select device and set maximum
integration time. Used in auto-
range and device will never
integrate longer
:SENSe:ADJMin
1: device
2: min. adjust
level
Select device and set desired
detect level level in auto-range
:SENSe:SHUTter
1: device
2: close/open
Select device, 2D, spectrometer
or luminance meter and set
autorange close or open
7.4.1
:SENSe:INT
Set integration time of the selected device. Example:
Write
:SENSe:INT 0,10000
ASCII CRLF
Read
Not Applicable
Integration time of the 2D imager is set to 10.000 us (10ms).
Device
Description
Min. int time [us]
Max. int time [us]
0
2D imager
1.000
10.000.000
1
Spectrometer
1.500
10.000.000
2
Luminance meter
1.000
5.000.000
Note: not all Atlas configurations have the luminance/flicker unit incorporated.
7.4.2
:SENSe:AVERage
Set average of the selected device. Example:
Write
:SENSe:AVERage 1,10
ASCII CRLF
Read
Not Applicable
Set average of spectrometer to 10.
Device
Description
Min. average
Max. average
0
2D imager
1
16
1
Spectrometer
1
20
2
Luminance meter
1
20
Note: not all Atlas configurations have the luminance/flicker unit incorporated.