7.3 Display settings commands
In the project settings commands the characteristics of the display and the
settings for the Capture display algorithm are set.
Command
Parameters
Purpose
:SENSe:FREQuency
1: Frequency
Set display refresh frequency
:SENSe:DS
1: horizontal
2: vertical
Set display size in mm
:SENSe:DR
1: horizontal
2: vertical
Set real display resolution in
pixels
:SENSe:VR
1: horizontal
2: vertical
Set virtual display resolution in
virtual pixels
:SENSe:AR
1: aspect
ratio
Set aspect ratio of display
:SENSe:WD
1: height
Set height(working distance) of
atlas to display in mm
7.3.1
:SENSe:FREQ
Set frequency of the Device Under Test. Many samples which need to be
tested have their own refresh frequency (display 60Hz for example) or a
PWM frequency (LED panels, 1kHz). In order to function properly our
auto-range algorithm works with integration steps of 1/freq. If this setting is
not set properly the measurement can be unstable.
Example:
Write
:SENSe:FREQ 60
ASCII CRLF
Read
Not Applicable
Frequency for the 2D imager is set to 60Hz.
Minimum frequency
Maximum frequency
10
100
In the case of high frequencies one should set the frequency to a multitude
of frames. For example if one has a LED panel of 1kHz the setting should
be set to 100Hz. In this case the auto-range function will work with 10
frames of the LED panel signal enhancing stability greatly.
If frequencies are very high the minimum integration time of the device
cannot be achieved. This is out of its range. Therefore you need to work
with more frames.