Admag SE100DJ/EJ User Manual Download Page 4

IM 1E10D0-01E

1-2

1.  INTRODUCTION

 Warranty

• The guaranteed term of this instrument is described

in the quotation. We repair the damages that
occurred during the guaranteed term for free.

• Please contact with our sales office when this

instrument is damaged.

• If the instrument has trouble, please inform us

model code, serial number, and concrete substances
or situations. It is preferable to be attached a outline
or data.

• We decide after the examination if free repair is

available or not.

• Please consent to the followings for causes of

damages that are not available as free repair, even if
it occured during the guaranteed term.

A: Unsuitable or insufficient maintenance by the

customer.

B: The handling, using, or storage that ignore the

design and specifications of the instrument.

C: Unsuitable location that ignore the description in

this manual.

D: Remaking or repair by a person except whom we

entrust.

E: Unsuitable removing after delivered.
F: A natural disaster (ex. a fire, earthquake, storm and

flood, thunderbolt) and external causes.

 For Safety Using

For safety using the instrument, please give attention
mentioned below.

WARNING

(1) Installation
• The instrument must be installed by expert

engineer or skilled personnel.  The procedures
described about INSTALLATION are not
permitted for operators.

• The Magnetic Flow Tube is a heavy instrument.

Please give attention to prevent that persons
are injured by carrying or installing. It is prefer-
able for carrying the instrument to use a cart
and be done by two or more persons.

• In case of high process temperature, care

should be taken not to burn yourself because
the surface of body and case reach a high
temperature.

• When removing the instrument from hazardous

processes, avoid contact with the fluid and the
interior of the flow tube.

• All installation shall comply with local install-

ation requirement and local electrical code.

(2) Wiring
• The instrument must be installed by expert

engineer or skilled personnel.  The procedures
described about WIRING are not permitted for
operators.

• Please confirm voltages between the power

supply and the instrument before connecting
the power cables.  And also, please confirm
that the cables are not powered before con-
necting.

• The protective grounding must be connected to

the terminal 

 in order to avoid personal shock

hazard.

(3) Operation
• Wait 10 min. after power is turned off, before

opening the covers.

(4) Maintenance
• Please do not carry out except being written to

a maintenance descriptions.  When these
procedures are needed, please contact to
nearest YOKOGAWA office.

• Care should be taken to prevent the build up of

drift, dust or other material on the display glass
and data plate.  In case of its maintenance, soft
and dry cloth is used.

(5) Explosion Protected Type Instrument
• For explosion proof type instrument, the de-

scription in Chapter 6 “EXPLOSION PRO-
TECTED TYPE INSTRUMENT” is prior to the
other description in this user's manual.

• Only trained persons use this instrument in the

industiral location.

• The protective grounding 

 must be connected

to a suitable IS grounding system.

• Take care not to generate mechanical spark

when access to the instrument and peripheral
devices in hazardous locations.

(6) The Instrument in Compliance with PED
• For the instrument in compliance with PED, the

description in Chapter 7 “PRESSURE EQUIP-
MENT DIRECTIVE” is prior to the other de-
scription in this User’s Manual.

Summary of Contents for SE100DJ/EJ

Page 1: ...User s Manual Y okogawa Electric Corporation Models SE100DJ EJ SE200DJ EJ and SE300DJ EJ Magnetic Flow Tube IM 1E10D0 01E IM 1E10D0 01E 8th Edition...

Page 2: ...recautions 3 3 3 2 2 Flow Tube Piping 3 4 3 3 Mounting 3 4 3 3 1 Nominal Diameter 15mm 0 5 to 40mm 1 5 Wafer Type 3 4 3 3 2 Nominal Diameter 50 mm 2 to 200 mm 8 Wafer Type 3 6 3 3 3 Nominal Diameter 1...

Page 3: ...recautions The following general safety precautions must be observed during all phases of operation service and repair of this instrument Failure to comply with these precautions or with specific WARN...

Page 4: ...self because the surface of body and case reach a high temperature When removing the instrument from hazardous processes avoid contact with the fluid and the interior of the flow tube All installation...

Page 5: ...e Precautions In case the instrument is expected to be stored over a long term please give attention to the followings The instrument should be stored in its original packing condition The storage loc...

Page 6: ...ucer pipe Expander pipe Tee 90 degrees bent Various types of valves 5 D or more 2 D or more 2 D or more 2 D or more 10 D or more 10 D or more 5 D or more 5 D or more Do not care to 0 D Do not care to...

Page 7: ...uid is always filled in the pipes The Vertical Mounting is effective for fluids that is easily separate or slurry settles within pipes In this case please flow a fluid from bottom to up GOOD GOOD BAD...

Page 8: ...prevent persons from injuring when it is handled 3 2 1 General Precautions 1 Precaution for Carrying The Magnetic Flow Tube is packed tightly When it is unpacked please give attention to prevent dama...

Page 9: ...elding scrap or wood chips may exist Please remove them by flushing piping before mounting the flow tube Slant Mis alignment F030202 EPS Figure 3 2 2 Slant and Mis alignment of Flow Tube Piping 3 3 Mo...

Page 10: ...pieces Please use appropriate bolts and nuts according to Process connection Horizontal Mounting Vertical Mounting F030301 2 EPS Figure 3 3 1 Mounting Procedure Size 15 mm 0 5 to 40 mm 1 5 Table 3 3 1...

Page 11: ...es two for each and lower adjacent holes of both flanges Refer to Figure 3 3 2 Please give attention to the nominal size and flange ratings of the centering devices Refer to Table 3 3 5 3 Positioning...

Page 12: ...has equal elasticity T030303 EPS Table 3 3 4 Tightening Torque Values for PVC Piping in N m kgf cm in lbf Size mm inch JIS 10K JIS 20K ANSI 150 ANSI 300 DIN PN10 50 2 80 3 100 4 150 6 200 8 8 2 84 73...

Page 13: ...termi nal box In case the measuring fluid flows against the arrow direction please refer to the section Reversing Flow Direction in the Instruc tion Manual of SE14 Magnetic Flow Converter Flange Flow...

Page 14: ...363 531 65 0 93 0 663 949 575 823 125 1 151 5 1280 1550 1112 1347 153 5 180 9 1570 1850 1365 1608 94 0 125 0 959 1275 832 1106 135 9 164 3 1390 1680 1208 1460 154 5 199 5 1580 2040 1373 1773 151 5 192...

Page 15: ...er if it is too short Please replace the short cable with the cable which is appropriate length wholly The all cable ends are to be provided with round crimp on terminal The signal cables must be rout...

Page 16: ...cables JIS C3401 or Polyvinyl chloride insulated and sheathed portable power cables JIS C3312 or equivalents Outer Diameter 6 5 to 12mm in diameter 10 5 to 11 5 mm for waterproof gland ECG ECU Nominal...

Page 17: ...terminal C and terminal A B Between terminal A and terminal B Between terminal EX1 and ground Between terminal EX2 and ground All insulation measurements must be performed with a 500V megger Insulati...

Page 18: ...converter Please refer to Instruction Manual of SE14 magnetic flow converter IMPORTANT Be sure to disconnect the cables at the terminals of the flow tube when checking 1 Excitation Coil Check Check th...

Page 19: ...and earth ring for optional code FRG Non asbestos joint sheet sheathed with fluoro resin PTFE between earth ring and process flange for optional code BSF Other gaskets between flow tube and earth rin...

Page 20: ...00 150 200 250 300 350 400 0 1909 0 5302 1 3572 2 1206 5 429 8 483 19 086 33 93 53 02 76 35 103 91 135 72 Size 6 361 17 671 45 23 70 68 180 95 282 74 636 1 1 130 9 1 767 1 2 544 6 3 463 4 523 SI Units...

Page 21: ...for 80 to 150mm because of same mating dimensions Note 2 Only ANSI1 2NPT electrical connection is available for FM or CSA explosion proof type JIS G1 2 electrical connection is not available for any...

Page 22: ...ing Treatment Degreased cleansing treatment Packing with desiccant Coating is changed to three layer coating Urethane coating on two layer Epoxy coating X2 K5 GOS Calibration Certificate for GOST Only...

Page 23: ...8 1 H H1 L 1 111 4 4 48 1 9 80 5 3 2 Lining Outside diameter Height Fluorocarbon PFA 86 3 4 D D D Ground Terminal SD2d eps 1 When no earth ring is selected the face to face length is shorter by approx...

Page 24: ...11 5 11 5 11 3 13 0 10 5 6 4 6 4 6 1 6 2 5 4 5 5 4 9 4 9 0 7 0 7 1 1 2 0 2 4 3 1 Flange outside dia Dia of holes Number of holes Thickness Except P T Earth ring D C 10 1 Lining Fluorocarbon PFA Fluor...

Page 25: ...ole pitch 22 5 22 5 22 5 22 5 22 5 22 5 22 5 22 5 22 5 45 45 45 150mm 6in 200mm 8in Flange Type Unit mm Approx inch Ground Terminal 80 5 3 2 48 1 9 86 3 4 111 4 4 20 5 0 8 68 2 7 H1 H2 L 1 T 2 H C D N...

Page 26: ...10 16 400 16 22 0 9 26 1 431 8 17 450 8 18 25 4 1 31 7 1 2 414 16 3 23 0 9 40 1 6 59 2 3 34 1 3 36 1 4 34 1 3 34 1 3 400 15 7 25 1 44 1 7 430 16 9 27 1 1 16 16 10 12 12 16 12 96 104 87 88 87 87 91 SE3...

Page 27: ...nstallation WARNING All wiring shall comply with local installation requirements and local electrical code In hazadous locations the cable entry devices shall be of a certified flameproof type suitabl...

Page 28: ...ure 70 C 85 C 120 C 150 C Minimum Process Temperature 40 C 40 C 40 C 40 C Enclosure NEMA 4X Ambient Temperature 20 to 60 C Maximum Working Pressure 4MPa SE115E to SE205E 2MPa SE208E to SE220E Flange r...

Page 29: ...of Certification 6 4 SAA 1 Technical Data SA Certificate No AUS Ex 3764X Type of Protection Ex d m ia II C T6 T3 Enclosure Type IP67 Excitation Circuit 41Vmax 6 6 25Hz Electrode Circuit Um 250V ac dc...

Page 30: ...02D E are not attached CE mark of PED because they do not come under CE marking of PED 2 Installation WARNING Please tighten the bolts for piping joint according to the prescribed torque values Please...

Reviews: