Admag SE100DJ/EJ User Manual Download Page 28

IM 1E10D0-01E

6-2

6.  EXPLOSION PROTECTED TYPE INSTRUMENT

ENCLOSURE : Enclosure protection number
LININGL : Material of liner
TAG NO. : Tag number
SIZE : Nominal size of apparatus
METER FACTOR : Sensor constant number of apparatus
ELECTRODE : Material of electrodes
FLUIDPRESS : Fluid pressure of apparatus
FLUIDTEMP. : Fluid temperature of apparatus
Tamb : Ambient temperature
No. : Manufacturing serial number
CE : CE marking
ExII 2G : Group II Catergory 2 Gas atmosphere
KEMA No. : KEMA 98ATEX3230 : Certificate number
EExdem[ia]IICT6...T3 : Protection type and temp. class
ELECTRODE CIRCUIT Um : Voltage of electrode circuit
IM : User's Manual

WARNING

 : Warning to apparatus

YOKOGAWA 

 

 TOKYO 180-8750 JAPAN :

Name and address of manufacturer

*1) The third figure from the last shows the last one

figure of the year of production.  For example, the
year of production of the product engraved as
follows is year 1998.

No. F261GA091 813

Produced in 1998

*2) The identification number of the notified body :

   0344

*3) The product-producing country

6.2 FM

(1) Technical Data

Explosionproof for Class I Division 1 Groups A, B,

C & D. Dust-ignitionproof for Class II/III Division
1 Groups E, F & G. Intrinsically safe (electrode) for
Class I Division 1 Groups A, B, C & D

Electrode circuit Vmax : 250V ac/dc

Temperature

Code

T6
T5
T4
T3

Maximum Ambient

Temperature

+60˚C
+60˚C
+60˚C
+60˚C

Maximum Process

Temperature

+70˚C
+85˚C

+120˚C
+150˚C

Minimum Process

Temperature

-40˚C
-40˚C
-40˚C
-40˚C

Enclosure : NEMA 4X
Ambient Temperature : -20 to +60

°

C

Maximum Working Pressure : 4MPa(SE115E to

SE205E), 2MPa(SE208E to SE220E)

    Flange rating should be also considered.

(2) Wiring

WARNING

• All wiring shall comply with national electrical

code ANSI/NFPA 70 and local electrical code.

• There is no need of conduit seal for both of

Division 1 and Division 2 hazardous locations
because this product is sealed at factory.

(3) Operation

WARNING

• OPEN CIRCUIT BEFORE REMOVING

COVER.  INSTALL IN ACCORDANCE WITH
THE INSTRUCTION MANNUAL IM1E10D0-
01E.

• Take care not to generate mechanical spark

when access to the instrument and peripheral
devices in hazardous locations.

(4) Maintenance and Repair

WARNING

The instrument modification or parts replacement
by other than authorized representative of
Yokogawa Electric Corporation is prohibited and
will void the approval of Factory Mutual
Research Corporation.

6.3 CSA

(1) Technical Data

   Class I, Groups B, C and D; Class II, Groups E, F

and G; Class III; Encl Type 4X

Electrodes: Intringically Safe, Ex ia, Class I, Groups

A, B, C and D

When installed in Class I, Division 2 locations -

Seals No Required.

Electrode Circuit Vmax : 250V ac/dc

Temperature

Code

T6
T5
T4
T3

Maximum Ambient

Temperature

+60˚C
+60˚C
+60˚C
+60˚C

Maximum Process

Temperature

+70˚C
+85˚C

+120˚C
+150˚C

Minimum Process

Temperature

-40˚C
-40˚C
-40˚C
-40˚C

Ambient Temperature : -20 to +60

°

C

    Maximum Working Pressure : 4MPa(SE115E to

SE205E), 2MPa(SE208E to SE220E)

    Flange rating should be also considered.

Summary of Contents for SE100DJ/EJ

Page 1: ...User s Manual Y okogawa Electric Corporation Models SE100DJ EJ SE200DJ EJ and SE300DJ EJ Magnetic Flow Tube IM 1E10D0 01E IM 1E10D0 01E 8th Edition...

Page 2: ...recautions 3 3 3 2 2 Flow Tube Piping 3 4 3 3 Mounting 3 4 3 3 1 Nominal Diameter 15mm 0 5 to 40mm 1 5 Wafer Type 3 4 3 3 2 Nominal Diameter 50 mm 2 to 200 mm 8 Wafer Type 3 6 3 3 3 Nominal Diameter 1...

Page 3: ...recautions The following general safety precautions must be observed during all phases of operation service and repair of this instrument Failure to comply with these precautions or with specific WARN...

Page 4: ...self because the surface of body and case reach a high temperature When removing the instrument from hazardous processes avoid contact with the fluid and the interior of the flow tube All installation...

Page 5: ...e Precautions In case the instrument is expected to be stored over a long term please give attention to the followings The instrument should be stored in its original packing condition The storage loc...

Page 6: ...ucer pipe Expander pipe Tee 90 degrees bent Various types of valves 5 D or more 2 D or more 2 D or more 2 D or more 10 D or more 10 D or more 5 D or more 5 D or more Do not care to 0 D Do not care to...

Page 7: ...uid is always filled in the pipes The Vertical Mounting is effective for fluids that is easily separate or slurry settles within pipes In this case please flow a fluid from bottom to up GOOD GOOD BAD...

Page 8: ...prevent persons from injuring when it is handled 3 2 1 General Precautions 1 Precaution for Carrying The Magnetic Flow Tube is packed tightly When it is unpacked please give attention to prevent dama...

Page 9: ...elding scrap or wood chips may exist Please remove them by flushing piping before mounting the flow tube Slant Mis alignment F030202 EPS Figure 3 2 2 Slant and Mis alignment of Flow Tube Piping 3 3 Mo...

Page 10: ...pieces Please use appropriate bolts and nuts according to Process connection Horizontal Mounting Vertical Mounting F030301 2 EPS Figure 3 3 1 Mounting Procedure Size 15 mm 0 5 to 40 mm 1 5 Table 3 3 1...

Page 11: ...es two for each and lower adjacent holes of both flanges Refer to Figure 3 3 2 Please give attention to the nominal size and flange ratings of the centering devices Refer to Table 3 3 5 3 Positioning...

Page 12: ...has equal elasticity T030303 EPS Table 3 3 4 Tightening Torque Values for PVC Piping in N m kgf cm in lbf Size mm inch JIS 10K JIS 20K ANSI 150 ANSI 300 DIN PN10 50 2 80 3 100 4 150 6 200 8 8 2 84 73...

Page 13: ...termi nal box In case the measuring fluid flows against the arrow direction please refer to the section Reversing Flow Direction in the Instruc tion Manual of SE14 Magnetic Flow Converter Flange Flow...

Page 14: ...363 531 65 0 93 0 663 949 575 823 125 1 151 5 1280 1550 1112 1347 153 5 180 9 1570 1850 1365 1608 94 0 125 0 959 1275 832 1106 135 9 164 3 1390 1680 1208 1460 154 5 199 5 1580 2040 1373 1773 151 5 192...

Page 15: ...er if it is too short Please replace the short cable with the cable which is appropriate length wholly The all cable ends are to be provided with round crimp on terminal The signal cables must be rout...

Page 16: ...cables JIS C3401 or Polyvinyl chloride insulated and sheathed portable power cables JIS C3312 or equivalents Outer Diameter 6 5 to 12mm in diameter 10 5 to 11 5 mm for waterproof gland ECG ECU Nominal...

Page 17: ...terminal C and terminal A B Between terminal A and terminal B Between terminal EX1 and ground Between terminal EX2 and ground All insulation measurements must be performed with a 500V megger Insulati...

Page 18: ...converter Please refer to Instruction Manual of SE14 magnetic flow converter IMPORTANT Be sure to disconnect the cables at the terminals of the flow tube when checking 1 Excitation Coil Check Check th...

Page 19: ...and earth ring for optional code FRG Non asbestos joint sheet sheathed with fluoro resin PTFE between earth ring and process flange for optional code BSF Other gaskets between flow tube and earth rin...

Page 20: ...00 150 200 250 300 350 400 0 1909 0 5302 1 3572 2 1206 5 429 8 483 19 086 33 93 53 02 76 35 103 91 135 72 Size 6 361 17 671 45 23 70 68 180 95 282 74 636 1 1 130 9 1 767 1 2 544 6 3 463 4 523 SI Units...

Page 21: ...for 80 to 150mm because of same mating dimensions Note 2 Only ANSI1 2NPT electrical connection is available for FM or CSA explosion proof type JIS G1 2 electrical connection is not available for any...

Page 22: ...ing Treatment Degreased cleansing treatment Packing with desiccant Coating is changed to three layer coating Urethane coating on two layer Epoxy coating X2 K5 GOS Calibration Certificate for GOST Only...

Page 23: ...8 1 H H1 L 1 111 4 4 48 1 9 80 5 3 2 Lining Outside diameter Height Fluorocarbon PFA 86 3 4 D D D Ground Terminal SD2d eps 1 When no earth ring is selected the face to face length is shorter by approx...

Page 24: ...11 5 11 5 11 3 13 0 10 5 6 4 6 4 6 1 6 2 5 4 5 5 4 9 4 9 0 7 0 7 1 1 2 0 2 4 3 1 Flange outside dia Dia of holes Number of holes Thickness Except P T Earth ring D C 10 1 Lining Fluorocarbon PFA Fluor...

Page 25: ...ole pitch 22 5 22 5 22 5 22 5 22 5 22 5 22 5 22 5 22 5 45 45 45 150mm 6in 200mm 8in Flange Type Unit mm Approx inch Ground Terminal 80 5 3 2 48 1 9 86 3 4 111 4 4 20 5 0 8 68 2 7 H1 H2 L 1 T 2 H C D N...

Page 26: ...10 16 400 16 22 0 9 26 1 431 8 17 450 8 18 25 4 1 31 7 1 2 414 16 3 23 0 9 40 1 6 59 2 3 34 1 3 36 1 4 34 1 3 34 1 3 400 15 7 25 1 44 1 7 430 16 9 27 1 1 16 16 10 12 12 16 12 96 104 87 88 87 87 91 SE3...

Page 27: ...nstallation WARNING All wiring shall comply with local installation requirements and local electrical code In hazadous locations the cable entry devices shall be of a certified flameproof type suitabl...

Page 28: ...ure 70 C 85 C 120 C 150 C Minimum Process Temperature 40 C 40 C 40 C 40 C Enclosure NEMA 4X Ambient Temperature 20 to 60 C Maximum Working Pressure 4MPa SE115E to SE205E 2MPa SE208E to SE220E Flange r...

Page 29: ...of Certification 6 4 SAA 1 Technical Data SA Certificate No AUS Ex 3764X Type of Protection Ex d m ia II C T6 T3 Enclosure Type IP67 Excitation Circuit 41Vmax 6 6 25Hz Electrode Circuit Um 250V ac dc...

Page 30: ...02D E are not attached CE mark of PED because they do not come under CE marking of PED 2 Installation WARNING Please tighten the bolts for piping joint according to the prescribed torque values Please...

Reviews: