1. The center frequency should be changed as the frequency value at the first
harmonic wave.
l
Press [MARKER] → PEAK
2. Change the span as 50 MHz, and display the signal in the center of screen again.
l
Press [SPAN] → 50MHz
l
Press [MARKER] → PEAK
3. Set the attenuation as 0 dB
:
l
Press [AMPT] → Attenuation → 0dB
4. To recognize whether the harmonic distortion component is produced by spectrum
analyzer, firstly display the input signal in trace 2
:
l
Press [TRACE] → Trace 2
l
Press [TRACE] → Clear
5. Place a Delta mark in the harmonic distortion component of trace 2:
l
Press [MARKER] → peak
l
Press [MARKER] → Delta
Summary of Contents for DSA2016