AD2821A OPERATION MANUAL
44
G
Note
:
Primary parameter accuracy(A
e
)
C
:
A
e
= 0.3%(1+C
x
/C
max
+C
min
/C
x
)
L
:
A
e
= 0.3% (1+ L
x
/L
max
+ L
min
/L
x
)
Z
:
A
e
= 0.3% (1+ Z
x
/Z
max
+ Z
min
/Z
x
)
R
:
A
e
=
0.3%(1+ R
x
/R
max
+ R
min
/R
x
)
Max and Min values are as follows:
Parameter
Range Auto
C
max
80
µ
F/f
C
min
150pF/f
L
max
159H/f
L
min
0.32mH/f
Z
max
1M
Ω
Z
min
1.59
Ω
Here
:
Z
max
= R
max
; Z
min
= R
min
, Frequency unit: kHz.
Secondary parameter accuracy
D
e
= A
e
/3
when D
x
≤
0.1
D
e
= A
e
(1+D
x
)/3
when D
x
>
0.1
Q
e
=
e
x
e
x
D
Q
D
Q
×
×
±
1
when
Q
x
·
D
e
<1
Summary of Contents for AD2821A