3.2.2
Simulate a Transmitter
When the Calibrator is used as simulate a two-wire transmitter,it can adjust the output loop current by
adjusting the Calibrator. There must be a 12V-28 V loop power to supply. Connect the test leads as
shown in figure.
1) Press to step to SIMULATE mA
mode.
The display shows “SIMULATE”
with unit “mA”.
2
)
The general connection is that red is
positive and black is negative,the red test
wire connected to " mV/V " and the black
test wire connected to "COM" socket.
3
)
If the current over 24mA, the display will
show “-OL-”.
4
)
Press in and rotary the knob to step
up/down in different increments.
5
)
Press key , stepping up/down 25%
of full scale for output value.
6
)
Press key to select auto ramp output
type.
7
)
Press key to select “SpanCheck”
function.