Volt/mA Calibrator
Using
3.2 Output
3.2.1 SOURCE mA Mode (Source mA)
Use this mode to supply current to a passive circuit.
There must be a path which allows the current to flow between “LOOP” terminal and “mV/V”
terminal, otherwise it
will show overload (figure flashing) when you set an output value.
1
)
Press to step to SOURCE mA mode.
The display shows “SOURCE” with unit “mA”
2
)
The test leads connected as shown in the right figure, the red test wire connected to "LOOP" and the
black test wire connected to "mV/V" socket.
3
)
If the current over 24mA, the display will show “-OL-”.
4
)
Press in and rotary the knob to step up/down in different increments.
5
)
Press key , stepping up/down 25% of full scale for output value.
6
)
Press key to select auto ramp output type.
7
)
Press key to select “SpanCheck” function.
11