background image

ZW60-210531-Rev006-UM-en 

 

7

 

5.4  Battery status indicator 

The scales are intended for rechargeable battery and/or mains operation. The rechargeable 
battery life is up to 100 hours depending on the set parameters. 

When  the  rechargeable  battery  needs  to  be  recharged,  the  battery  symbol  appears  on  the 
display. The scale will operate for about 30 minutes. After this time, the scale will switch off 
automatically  to  protect  the  rechargeable  battery  from  damage.  If  the  scale  is  switched  on 
when the battery charge is low, “bAt Lo” will appear on the display every 5 minutes. 

To recharge the rechargeable battery, simply plug the power adapter into the scale and switch 
on the mains voltage. The scale does not need to be switched on. The rechargeable battery 
must be charged for 12 hours to its full capacity. 

The scales are delivered with partially charged rechargeable batteries. Since a rechargeable 
battery may get discharged in storage, it must be fully charged before the first use. 

LED display 

Green 

Battery is fully charged. 

Yellow 

The yellow light indicates that the rechargeable battery should be charged longer 
(preferably overnight).  

Battery is charging. 

red 

Battery almost discharged. Please charge immediately. Charging can take up to 
twelve hours. 

 

In case of improper use of the rechargeable battery or if the rechargeable battery is used for 
several years without replacement, it is possible that the rechargeable battery will no longer 
be fully charged. If the rechargeable battery life becomes unacceptable, contact your customer 
service representative or the manufacturer. 

6.  Operation 

6.1  Zeroing the scale 

When the  gross  weight  is  within ±2%  of zero,  a new  zero  point  is set  as  when the  scale  is 
switched on. When the gross weight exceeds ±2%, the tare function is performed. 

You can set the zero point at any time by pressing the ZERO button. All weighing and counting 
operations are performed from this point. Normally, it is only necessary if the weighing platform 
is unloaded. When the zero point is reached, the display shows “

Zero

”. 

Notice

: The scale has an auto zero function to account for small deviations or accumulation 

of items on the weighing platform. However, it is possible that you will need to press ZERO to 
reset the scale to zero if small weight quantities are displayed while the weighing platform is 
unloaded. 

6.2  Taring 

Press the TARE button to tare the scale. The displayed weight will be saved as a tare weight 
and the value will be subtracted from the display. Zero remains on the display. The display 
will show “

NET

”. 

When a product is added, only its weight is shown on the display. The scale can be tared for 
a second time if another product is to be added to the first one. This process can be repeated 
until the upper limit of the tare range is reached. 

Summary of Contents for ZW60-45

Page 1: ...OPERATING MANUAL Electronic table scales Control scale Model ZW60 45 ZW60 210531 Rev006 UM en ...

Page 2: ...roing the scale 7 6 2 Taring 7 6 3 Determining the weight 8 6 4 Weighing in percent 8 6 5 Piece counting 8 6 5 1 Method 1 Unknown reference weight 8 6 5 2 Method 2 Known reference weight 8 6 6 Presetting the maximum and minimum values for audible alarm 9 6 7 Totaliser 9 6 7 1 Manual totaliser 10 6 7 2 Automatic totaliser 10 7 Parameters 11 8 Care and maintenance 13 8 1 Cleaning 13 8 2 Error indica...

Page 3: ...between such units as kg g lbs and oz Power is supplied via a 230 V mains connection or a 6 V 4 Ah rechargeable battery 2 Delivery contents Included in the delivery are Scales Rechargeable battery integrated into the scale Weighing platform Power adapter User manual incl declaration of conformity 3 Safety instructions It is essential that you read understand and follow all instructions in these in...

Page 4: ...es 1 Carefully remove the delivery contents from the packaging 2 If necessary remove the transport lock before operating the scale The transport lock if any is located on the underside of the scale 3 Place the scale on a stable horizontal surface Do not place the scale on a shaky or vibrating table as this will cause distortion of measurement results 4 Do not place the scale near vibrating machine...

Page 5: ...on Skips the active digit one place to the right when entering values of adjustable parameters or other functions MODE Mode button Changes the units in weighing mode Changes display between unit weight counted quantity and total weight in counting mode Displays weight in percent mode The scale automatically returns to percent mode in three seconds Secondary function Skips the active digit one plac...

Page 6: ...weight value in weighing mode or the minimum quantity in counting mode If the applied weight or the applied quantity is smaller than the minimum value you will hear an audible signal To enter the value press and hold down this button for 3 seconds 0 9 Manual entry of values Numeric buttons from 0 to 9 are used to manually enter values such as unit weight or reference quantity C Clear manually ente...

Page 7: ...rgeable battery or if the rechargeable battery is used for several years without replacement it is possible that the rechargeable battery will no longer be fully charged If the rechargeable battery life becomes unacceptable contact your customer service representative or the manufacturer 6 Operation 6 1 Zeroing the scale When the gross weight is within 2 of zero a new zero point is set as when the...

Page 8: ...return to the weighing mode 6 5 Piece counting 6 5 1 Method 1 Unknown reference weight First possibly tare the empty container as described above and leave it on the weighing surface Press the SMPL button to start counting pieces Place a certain number of items to be weighed on the weighing surface You must know the exact number Enter this number using the numeric keypad Press the SMPL button agai...

Page 9: ...play shows OK The minimum value is set and the maximum value is zero The display shows OK as long as the weight is lower than the set minimum value When the applied weight exceeds the minimum value the display shows HIGH The minimum value is zero and the maximum value is set The display shows LOW as long as the applied weight is lower than the set maximum value When the applied weight exceeds the ...

Page 10: ...not upgrade able To clear the totaliser press the MC button The M sum function is enabled only if the display shows STABLE 6 7 2 Automatic totaliser The scale can be set so that the weight values on the display are automatically added to the values in the totalis er For this purpose parameter P6 Ato must be enabled setting Au on For more details see parameter settings When the automatic totaliser ...

Page 11: ...E button moves to the next lower setting Press the ZERO button again to save your entries Press the G N button to exit the menu without saving Please note that some functions are only reserved for use by authorised technicians according to the overview below PARA METER SUB PARA METER DISPLAY SETTING P0 CAL External calibration Reserved for use by authorised technicians only P1 LCA Linearity calibr...

Page 12: ...ptions Auto on off Auto Backlight is enabled when the weighing surface is loaded on Backlight is permanently enabled of Backlight is permanently disabled P12 bP BEEP Audible signal setting for checkweighing 0 Audible signal is disabled 1 Audible signal between two values is enabled 2 Audible signal beyond two values is enabled P13 ti Time Year date and time setting Navigate within the digits with ...

Page 13: ...tch on zero exceeds the permissible value typically 4 of the maximum weighing range when the scale is switched on or when the ZERO button is pressed The weighing platform has already been loaded when the scale has been switched on Weighing platform overload when the scale is set to zero Improper calibration of the scale Damaged load cell Damaged electronics Err 5 Keypad fault Improper use of the s...

Page 14: ...idence is provided that it has failed within 1 year from the date of sale proof of purchase is required due to lack of workmanship or defective material All moving parts such as batteries cables etc are excluded The use of accessories other than those approved by ADE especially batteries will void the warranty This warranty does not cover external normal wear and tear or damage caused by accident ...

Page 15: ... adapter Calibration With external weight Display 6 digit LCD digital displays 52 mm digit height Housing ABS plastic housing stainless steel weighing platform Weighing platform 365 x 235 mm Housing dimensions WxDxH 365 x 365 x 135 mm Net weight 5 1 kg Applications Multi purpose scale with simple piece counting function Functions Weighing piece counting totaliser freely programmable minimum and ma...

Page 16: ...rer This product has been manufactured in accordance with the harmonised European standards They conform to the regulations of the EC directives listed below EMC Directive 2014 30 EU Low voltage directive 2014 35 EU RoHS Directive 2011 65 EC in the currently valid versions This declaration will lose its validity if any modification is made to the scale without our consent Hamburg May 2021 ADE Germ...

Page 17: ...ZW60 210531 Rev006 UM en 17 ...

Page 18: ...18 ZW60 210531 Rev006 UM en ...

Page 19: ...ZW60 210531 Rev006 UM en 19 ...

Page 20: ...20 ZW60 210531 Rev006 UM en Händler ...

Reviews: