5.  Place the patient into the bed. The displayed weight is the weight of the patient.  

6.  If the weight of the patient should be ignored as well (e.g. for dialysis purposes), follow the instruction 

under the point “Tare function”. 

7.  If the weight of the bed including bed-clothes is already being stored follow the instruction under the 

point “Restore the preset tara value” 

Weighing 

Press  the  ON  key  for  at  least  3  seconds

 

to  turn  the  scales  on.  An  automatic  function  test  will  be  conducted 

afterwards. The test is done when the display shows zero. The scale is now ready for the weighing procedure. 
 
Resetting the scales back to zero: Wait until the standstill status indicator 

 (14) appears. By entering the 

ZERO

 

key (8), measuring inaccuracies after switching on the scale are being compensated. The zero key 

restarts the scale. The status indicator (6)             appears when the scale is being resetted. 

Tare function 

The  TARE  key  (9)  is  useful  for  many  weighing  operations.  Tare  might  be  useful  in  case

 

the  patient  is  being 

place on the scale/bed with items (e.g. clothing, bed clothing shoes or other items) which should be ignored by 
the scale.  Weigh these  items first. After the weight  is displayed, press the TARE key (9) to reset the display 
back to zero. The display changes to net mode (NET status indicator (5) is on) and shows zero. 
 
Note: Taring the bed with subsequent determination of the patient weight is not possible/practical, as the bed 

weight is subtracted from the maximum load of the scale and thus only a small weighing range is avail-
able. To determine the patient weight, proceed according to section "Before Weighing" on page 7. 

Entering manually tare - values 

 

The user is able to insert manually the gross weight, which is for example mentioned at the identification plate 
of the bed. Enter the appropriate tare value just by using the key pad under normal weighing mode. The value 
will be stored after use of the ENTER key (or TARE). The ADE-display is now ready for weighing and changes 
to net mode [NET status indicator (5) is on]. Enter the tare button to leave this mode. 

Preset tara 

Press and hold the PRESET TARE key

 

(10) for approx. 3 seconds. 

PRO PT X

 appears on the display. Select 

the memory location 1-10. Now enter the TARE VALUE which you would like to be ignored by the scale and 
confirm same with  ENTER. The value  is displayed  as a  negative figure and will  be deducted  when a load is 
being placed on the scale. Enter the on/off, zero – button (8) to leave this mode. 

Restore the preset tara value 

Press the PRESET TARE key (10). "Pt  x" appears on the display. Enter a number from 0-9 (memory location 
for  the  tare  value).  If  no  new  number  will  be  inserted  to  change  the  already  displayed  value  within  five  sec-
onds, the unit will turn back to weighing mode and displays the last stored value as negative figure as above. 
Enter the on/off, zero – button (8) to leave this mode. 

Displaying the gross weight, net weight and tare value 

Press  the  "0"  key  to  change  between  the  net  weight  (gross  weight  minus  the  tared  value)  and  the  gross 
weight.  

Hold function 

The  determined  weight  could  be  continued  being  displayed  even  when  the  load  has  already  been  removed. 
That enables taking care of the patient first before reading the results. 
 
To  use  the  hold  function  the  scale  must  be  at  standstill.    The  standstill  status  indicator 

  (14)  is  on  and  

Hold

 appears at the digital display. Deactivate the HOLD function by pressing the key again. 

Summary of Contents for M600020

Page 1: ...Operating Manual Class III Approved Scales Electronic Bed Scale M600020 M600020 210322 Rev008 UM e...

Page 2: ...ional devices 5 Housing and connections 6 Safety 6 Settings 6 Setting date and time 7 Operation 7 Modes of operation 7 Before weighing 7 Weighing 8 Tare function 8 Hold function 8 Determining the Body...

Page 3: ...ry place Dismantle packaging and foil carefully and remove the transport securing devic es Place the weighing feet on a plain dry and level surface before the weighing procedure Scope of delivery 4 We...

Page 4: ...ging period the older they are If the battery performance drops considerably then it is time to replace the battery New batteries are supplied in partial charged condition Batteries achieve their full...

Page 5: ...able for CE approved class III scales 13 Number pad Figures 0 9 Clear Enter and Navigation functions Key 6 menu direction to the right Key 4 menu direction to the left Key 8 menu direction down Key 2...

Page 6: ...positioning and operation follow the instructions of this operation manual Portable and mobile HF communication systems may influence electronic medical devices Make sure that the mains cable does no...

Page 7: ...Confirm with ENTER and date time are set Operation Modes of operation Weighing mode The weighing mode is of course the standard mode of your scale After initial configuration and sealing by attaching...

Page 8: ...tion Before Weighing on page 7 Entering manually tare values The user is able to insert manually the gross weight which is for example mentioned at the identification plate of the bed Enter the approp...

Page 9: ...13 to send the data to the serial interface Note Printing is not possible if the net weight is 0 The print function only produces one print out A reset is required to print out another copy Care serv...

Page 10: ...very applying to defects oc curred due to poor material or workmanship presentation of invoice is necessary All moveable parts bat teries cables mains units rechargeable batteries etc are excluded War...

Page 11: ...t Full duplex RS 232 38400 19200 9600 4800 2400 1200 bps Printer Port Output only RS 232 or active 20 mA current loop 9600 4800 2400 1200 bps Both Ports 7 or 8 data bits even odd or no parity Operator...

Page 12: ...conformity by the manufacturer This product has been manufactured in accordance with the harmonised European standards They con form to the regulations of the EC directives listed below NAWI Directive...

Page 13: ...M600020 210322 Rev008 UM e 13...

Page 14: ...14 M600020 210322 Rev008 UM e...

Page 15: ...M600020 210322 Rev008 UM e 15...

Page 16: ...16 M600020 210322 Rev008 UM e Dealer...

Reviews: