ADE M50010 Operating Manual Download Page 6

5  

050100-Rev000-M50010 

 

Use of the scale 

 

Weighing 

 
Press 

ON

 key for min. 3 seconds to power the scale. After start up an automatic function test is done 

which ends in switching into `0.00´ position in kg weighing mode. The scale is now ready to weigh. In 
case of any irregularities in weighing procedure the display will show the error indication Err xxx. 
In this case go to chapter Cleaning, Maintenance and Re-approval, see point Service. 
After `0.00´ is displayed properly the scale can be loaded. Make sure there are no movements during 
weighing procedure.  
The correct weight can be read on the display. 
 

Body Mass Index (BMI) 

 
The Body Mass Index is the ration between height and weight squared. The BMI is meanwhile 
accepted world-wide – also by WHO (World Health Organization) – helping to evaluate the state of 
nutrition as well as the state of health of a person. The result is a tolerance value. 
 
As the exact body length is one essential value we recommend to measure the patient by using in 
addition to your ADE scale the ADE height measuring rod model MZ 10019. 
 
Start the scale as already described and take the weight.. Then press the 

BMI 

key and use the 

numeric key to indicate the height of the patient in cm. Confirm by using the 

ENTER

 key. The BMI is 

calculated and displayed. 
 

Evaluation of the Body Mass Index 

 
BMI below 18 

  The patient is not showing enough weight. Control nutrition to put on weight. 

BMI between 18 and 20 

  The patient is slightly under weighed. If he feels well there is no need for a change.. 

BMI between 20 and 25 

  The patient is in the range of normal weight. 

BMI between 25 and 30 

  The patient is slightly overweighed. A reduction of weight is indicated or, is necessary if the 

state of health is already affected (e.g. diabetes, high end blood pressure, heart diseases). 

BMI over 30 

  Reduction of weight is imperative. Also to visit a doctor is highly recommended. 

 

Tare functions 

 
The 

SEMI AUTO TARE

 key is used for weighing patients in different types of wheelchair models (for 

example in their own wheelchair) which means the weight of the wheelchair is unknown. In this case 
weigh empty wheelchair first. Activate the SEMI AUTO TARE key and the weight of the wheelchair is 
stored, the indication in the display set back to zero. Is the patient now weighed sitting in his 
wheelchair, only the weight of the patient is indicated. 
 
By using the 

Preset TARA 

key any requested value can be put under 

tare

 by manually keying in this 

one tare value using the numeric keys. Confirmation is done by the 

ENTER

 key and can be re-

activated on request. This function is useful whenever patients are usually weighed in the same 
wheelchair and therefore the value to be tared is a fixed one. 

Summary of Contents for M50010

Page 1: ...Operating Manual Electronic Wheelchair Scale class III approved M50010 ADE GmbH Co Hammer Steindamm 27 29 22089 Hamburg GERMANY 49 40 432776 0 Fax 49 40 432776 10 E mail info ade hamburg de...

Page 2: ...d scales are imperative 4 Connecting additional equipment 4 Quality check 4 EC type approval 4 Use of the scale 5 Weighing 5 Body Mass Index BMI 5 Evaluation of the Body Mass Index 5 Tare functions 5...

Page 3: ...y to be tared HOLD To fix the weighing result BMI To evaluate the BMI Body Mass Index SEMI AUTO TARE To tare a value automatically UNITS To change the weighing unit CLEAR Clear indication ENTER Confir...

Page 4: ...Manual including Certificate of Conformity Attention Carefully check identity of the serial numbers on the platform and the indicator which is mandatory for medical use of the scale Power Supply of t...

Page 5: ...pean Directive 93 42 EEC for medical devices and declares the conformity as per this Directive Due to the certification as per ISO 9001 2000 the high standards for Research Development Manufacturing S...

Page 6: ...BMI is calculated and displayed Evaluation of the Body Mass Index BMI below 18 The patient is not showing enough weight Control nutrition to put on weight BMI between 18 and 20 The patient is slightly...

Page 7: ...Display indication Lo Recharge the power pack by connecting the mains adapter to the display Display indication ErrL Check correct positioning of the scale see chapter installation use of scale and m...