Installation and use of scale 

 

 
Your ADE Electronic Wheelchair Scale is a class III approved scale which has been manufactured for 
medical use in hospitals, by doctors or medical specialists as well as in rehabilitation centres or 
nursing homes. 

 
The scale is used to evaluate the weight of patients sitting in a wheelchair, able to walk on their 
own or by using any kind of walking aid or lying on an emergency litter. Due to its high 
capacity the scale is also used for adipose patients. 

 

The scale is a precision instrument for medical use. 
The weighing result is used as an essential basis for medical application decided by doctors or 
medical personnel in hospitals, nursing homes, rehabilitation- and dialysis centres. 

 
In order to achieve the best possible results kindly read and follow the Operating Manual in all 
respects. 
 

Check carton contents 

 

Platform Scale equipped with 1 ramp 

Indicator and cable with jag connector system 

- Mains 

adapter 

Operating Manual including Certificate of Conformity 

 

Attention 

Carefully check identity of the serial numbers on the platform and the indicator which is 
mandatory for medical use of the scale. 

 
Power Supply of the scale 

 
By mains adapter (12V 300mA) and rechargeable batteries (power-pack).  
 

The battery pack is already installed in the display. 

Charge the power-pack by using the mains adapter and connect the jack firmly to the display. 

Complete first charging of power-pack needs 24 hours. 

If scale is used on batteries and display shows 

Lo 

power-pack needs to be recharged. 

Please make sure the local voltage supply corresponds to the indication printed on the 
adapter. 

 

Positioning of the scale 

 

Place the scale on a level surface. 

Make sure all four feet underneath the platform have direct contact to the floor. 

The spirit level of the scale (next to the handle) is an additional tool to make sure the scale is 
perfectly positioned and offers best results when the air bubble is centred. 

Connect the display to the platform by using the jag connector. 

 

The alignment of the platform must be checked and corrected if necessary after storage or 
every time the location of the scale was changed. 

Summary of Contents for M50010

Page 1: ...Operating Manual Electronic Wheelchair Scale class III approved M50010 ADE GmbH Co Hammer Steindamm 27 29 22089 Hamburg GERMANY 49 40 432776 0 Fax 49 40 432776 10 E mail info ade hamburg de...

Page 2: ...d scales are imperative 4 Connecting additional equipment 4 Quality check 4 EC type approval 4 Use of the scale 5 Weighing 5 Body Mass Index BMI 5 Evaluation of the Body Mass Index 5 Tare functions 5...

Page 3: ...y to be tared HOLD To fix the weighing result BMI To evaluate the BMI Body Mass Index SEMI AUTO TARE To tare a value automatically UNITS To change the weighing unit CLEAR Clear indication ENTER Confir...

Page 4: ...Manual including Certificate of Conformity Attention Carefully check identity of the serial numbers on the platform and the indicator which is mandatory for medical use of the scale Power Supply of t...

Page 5: ...pean Directive 93 42 EEC for medical devices and declares the conformity as per this Directive Due to the certification as per ISO 9001 2000 the high standards for Research Development Manufacturing S...

Page 6: ...BMI is calculated and displayed Evaluation of the Body Mass Index BMI below 18 The patient is not showing enough weight Control nutrition to put on weight BMI between 18 and 20 The patient is slightly...

Page 7: ...Display indication Lo Recharge the power pack by connecting the mains adapter to the display Display indication ErrL Check correct positioning of the scale see chapter installation use of scale and m...

Page 8: ...or any warranty claim customer service and spare parts Re approval For re approval of the scale as per the national directives governing the standards for non automatic scales as well as for medical d...

Page 9: ...i folgende direktiver 90 384 EEC om ikke automatiske vekter 93 42 EEC om medisinske proukter og den europeiske standarden DIN EN 45501 FIN on tyyppihyv ksynt todistuksessa kuvatun raken nemallin muka...

Page 10: ...9 050100 Rev000 M50010...

Page 11: ...050100 Rev000 M50010 10...

Page 12: ...11 050100 Rev000 M50010...

Reviews: