ADE M403660 Operating Manual Download Page 7

M403660-211129-Rev006-UM-e 

 

Correct weighing 

Enter 

ON/OFF

 to get into the normal operating mode of the scale with no load on the chair scale. 

An automatic function test will be performed afterwards and the zero setting of the scale. When 
`0.0´appears you are in the kg weighing mode. Additionally the display shows the symbol `

´ 

to confirm stability of the zero setting. 

The scale is now ready to be used. 

Help the patient into the chair scale and ensure that he is not moving. 

The weight could be only determined when there is no contact of the feet to the floor. For this 
reason ensure that the patient’s feet rest on the footrest. 

The utmost accuracy has been achieved. after the weight is being displayed and the symbol `



´ 

appears  

To turn off the scale press 

ON/OFF

 for a period of minimum 3 seconds. Using the scale on mains 

it may remain permanently switched on or can be shut off manually by pressing key. 

 

How to change the unit  

The scale offers the possibility to select between the weighing mode kg and lb. For changing the mode 
press the 

UNIT 

key. Please check weighing mode after switching on to get proper result. 

 

Tare function 

The tare function allows ignoring a fixed weight (e.g. clothes) that will be deducted after each weighing 
procedure. Proceed as follows: 

Press the start key with no load on the scale. 

Place the additional weight of the object(s), which needs to be ignored, on the scale and store 
the result by pressing the 

TARE

 key. 

The display is being reset to zero. The scale is now in NET weighing mode which is also 
indicated on the display. 

Remove now the objects from the platform. The ignored weight is being displayed with a 
minus. 

Place the Patient and the additional weight (e.g. clothes) on the scale. 

The scale determines the weight of the patient without the additional weight. You may now 
use the scale for as many weighing procedures and the stored value will always be deducted 
as long as the scale is not turned off. 

Enter the tare button again to get of the tare mode. 

The tare value is likewise erased by the scale until pressing the tare button or being switched 
off. 

 

Weighing with preset tare values 

To input a pre set tare value press down the tare button more than 3 seconds. NET will appear at the 
screen. Follow the same steps as before and place a load on the platform. The last stored tare value 
will appear at the display and ignored by the scale. 

 

Summary of Contents for M403660

Page 1: ...Operating Manual International Range Electronic Chair Scales M403660 M403660 211129 Rev006 UM e...

Page 2: ...scription 4 Key pad 5 Operation 5 Positioning of patient 5 How to change the unit 6 Tare function 6 Weighing with preset tare values 6 HOLD function 7 BMI Body Mass Index 7 Evaluation of the Body Mass...

Page 3: ...uctions in this operating manual the country specific installation standards as well as the applicable safety conditions and accident prevention regulations in all aspects Operate the scale only with...

Page 4: ...rgeable batteries Use the AC adapter to load the rechargeable batteries The imprinted voltage value at the mains adapter must match the local mains voltage When the battery is being charged for the fi...

Page 5: ...thin the zero level Standstill status indicator 14 The scale is at a standstill meaning the weight is stable and no movement is on the platform This is a mandatory precondition for functions such as r...

Page 6: ...witch between the numbers downwards In BMI mode use this key to switch between the numbers upwards Enter button to confirm the value Operation Positioning of patient Please ensure for security reasons...

Page 7: ...hing on to get proper result Tare function The tare function allows ignoring a fixed weight e g clothes that will be deducted after each weighing procedure Proceed as follows Press the start key with...

Page 8: ...w buttons to enter the height of the patient Confirm the height with the enter button The BMI is now being displayed To correct the given height figures press the BMI button once again and the display...

Page 9: ...thorized personnel Weight reading exceeds load limit Reduce load on scale until weight value can be displayed Weight reading below load limit Service by authorized personnel required Configuration or...

Page 10: ...elivery applying to defects occurred due to poor material or workmanship presentation of invoice is necessary All moveable parts batteries cables mains units rechargeable batteries etc are excluded Wa...

Page 11: ...dapter AC Adapter 6 9VDC 500mA Central negative 4xAA NIMH rechargeable batteries pack Display 6 digit 7 segment 1 25mm LCD with 16 annunciators and blue backlight Working temperature 10 C to 40 C Stor...

Page 12: ...ermany 49 40 432 776 0 49 40 432 776 10 info ade germany de www ade germany de Declaration of conformity by the manufacturer These scales have been manufactured in accordance with the harmonised Europ...

Page 13: ...M403660 211129 Rev006 UM e 12...

Page 14: ...13 M403660 211129 Rev006 UM e...

Page 15: ...M403660 211129 Rev006 UM e 14...

Page 16: ...15 M403660 211129 Rev006 UM e authorised dealer...

Reviews: