ADE M40010 Operating Manual Download Page 4

 

3  

050100-Rev000-M40010 

 

Installation and use of scale 

 

Your ADE Electronic Chair Scale is a class III approved scale which has been manufactured for 
medical use in hospitals, by doctors or medical specialists. 
 

The scale is used to evaluate the weight of patients who can only be weighed seated. 
Due to the foldable arm- and footrests it is necessary that the patient is supervised by 
trained personnel and is not left unattended at any time. Make sure the brake system is 
activated while transferring the patient into the seat. 
The scale is a precision instrument for medical use. 
The weighing result is used as an essential basis for medical application decided by doctors or 
medical personnel in hospitals, nursing homes, rehabilitation- and dialysis centres. 
 

In order to achieve the best possible results kindly read and follow the Operating Manual in all 
respects. 
 

Check carton contents 

 

Chair Scale, fully assembled 

- Mains 

adapter 

Operating Manual including Certificate of Conformity 

 

Power Supply of the scale 

 
By mains adapter (12V 300mA) and rechargeable batteries (power-pack).  
 

The battery pack is already installed in the display. 

Charge the power-pack by using the mains adapter and connect the jack firmly to the display. 

Complete first charging of power-pack needs 24 hours. 

If scale is used on batteries and display shows 

Lo 

power-pack needs to be recharged. 

Please make sure the local voltage supply corresponds to the indication printed on the 
adapter. 

 

Positioning of the scale 

 
Place the scale on a level surface. 
For proper alignment please use the two knurled nuts situated on the back wheels of the scale. 
The spirit level of the scale is an additional tool to make sure the scale is correctly aligned. In this case 
the air bubble is centred. 

The alignment of the scale has to be checked and corrected every time the location of the scale is 
changed. 

 

Use of the scale where approved scales are imperative 

 
The scale corresponds to standards and requirements of the European Directives 90 90/384/EEC and 
93/42 EEC.  
 

Connecting additional equipment 

 
Conditions for connecting additional equipment are fulfilled via RS 232. 
 
Metrologic requirements for additional equipment like printer and PC have to be metrologically marked 
by 

on green background

.

 

Summary of Contents for M40010

Page 1: ...Operating Manual Electronic Chair Scale class III approved M40010 ADE GmbH Co Hammer Steindamm 27 29 D 22089 Hamburg Germany 49 40 432776 0 Fax 49 40 432776 10 E mail info ade hamburg de...

Page 2: ...re approved scales are imperative 3 Connecting additional equipment 3 Quality check 4 EC type approval 4 Use of the scale 4 Weighing 4 Body Mass Index BMI 5 Evaluation of the Body Mass Index 5 Tare fu...

Page 3: ...e tared HOLD To fix the weighing result BMI To evaluate the BMI Body Mess Index SEMI AUTO TARE To tare a value automatically UNITS To change the weighing unit kg lbs CLEAR Clear indication ENTER Confi...

Page 4: ...12V 300mA and rechargeable batteries power pack The battery pack is already installed in the display Charge the power pack by using the mains adapter and connect the jack firmly to the display Comple...

Page 5: ...with class III sign are calibrated in accordance with the high precision standards laid down in EC Directive 90 384 EEC Products carrying this symbol meet the requirements of Directive for non automat...

Page 6: ...abetes high end blood pressure heart diseases BMI over 30 Reduction of weight is imperative Also to visit a doctor is highly recommended Tare functions The SEMI AUTO TARE key is used for taring object...

Page 7: ...shows visible damages If mains adapter is not working properly After long storage in humid places In those cases contact your local dealer or service partner Storage and Transport Keep all original pa...

Page 8: ...approval becomes necessary IF the approval seal has been damaged AFTER repair of an approved scale OR after expiry of the validity of the first approval which is nationally regulated For further info...

Page 9: ...i folgende direktiver 90 384 EEC om ikke automatiske vekter 93 42 EEC om medisinske proukter og den europeiske standarden DIN EN 45501 FIN on tyyppihyv ksynt todistuksessa kuvatun raken nemallin muka...

Page 10: ...9 050100 Rev000 M40010...

Page 11: ...050100 Rev000 M40010 10...

Page 12: ...11 050100 Rev000 M40010...

Reviews: