52
Connect the gauge as shown in Figure 20 (dial pressure gauge)/21 (digital pressure gauge).
Figure 20 Connect a dial pressure gauge Figure 21 Connect a digital pressure gauge
3.
Manually set or use the step function (see Para. 3.2.1 "Auto step and manual step of pressure output") to output the
pressure of each calibration point.
4. When the pressure of each calibration point becomes stable and is displayed in green, record the readings of each
point.
Summary of Contents for 762
Page 1: ...762 Automated Hydraulic Pressure Calibrator...
Page 2: ......
Page 4: ......
Page 6: ......
Page 18: ......
Page 29: ...11 2 Installation 2 1 Features 2 1 1 Basic structure Figure 1 Basic structure...
Page 50: ...32 Figure 9 Current measurement...
Page 139: ...121...