2.
Set up
Select application in the function menu to enter the pressure leak test
Click
to enter the setting
Click the upper area of the interface to select the range: select the internal module range or external pressure
module for leak testing
Select unit: If the selected unit is not supported by the selected module, it will automatically restore to the last
setting
When there is a voltage-controlled external module A/B online, the internal voltage control option will appear,
indicating whether to use the external module to control the voltage
set point
When there is an external controllable voltage module online, the external voltage control module is selected,
and the internal control voltage is turned on, the external control voltage set point can be set;
Setpoints can be set directly when online internal modules are selected.
Must not exceed the currently selected range;
Use this setting as the leak detection pressure point;
The target value when performing the test.
Set dwell time
Select the internal range: After the pressure control is stable during the execution (the real-time pressure
value turns green display), switch to pressure measurement, and start the countdown waiting time.
Select an external pressure module: the waiting time will be counted down after the execution starts.
After the waiting time is over, enter the test time.