Figure 5- 7 Calibration of the VAISALA digital barometer
2. New task
Select the “digital barometer” in the test type list, and input the information of the barometer.
Table 5- 12 Task mode digital barometer
Item
Effective value
Description
Name
Letter, numeral, symbol
Name of the digital barometer
Serial number
Letter, numeral, symbol
Serial number of digital barometer
Model
Letter, numeral, symbol
Model of the digital barometer
Calibration range
Depends on the barometer
The input upper and lower limits can’t be equal.
Allowable error
0.01~100%
Allowable error of digital barometer
Rounding of reading
1, 0.1, 0.01
Rounding format of the reading