time, and then click the OK button
to start the test.
When selecting the manual execution, after the set point is stable, it is required to click the next point button
to collect data and then input the new set point.
To stop this test, click the back button
.
3. Save the result
Click the save button
to save the test result.
The following items need to be entered: Name and serial number, click OK button
to save the test results.
To exit without saving press the back button
.
To redo the test click the reset button
.
5.1.3 Pressure switch
Select the “pressure switch” icon
in the test type list to start the quick test.
1. Connection
Connect the switch as shown in Figure 3-20.
2. Start the test
Click the zeroing button
to reset the pressure control module;
Click the start button
to select to enter the execution interface;