Trouble Area
No light output
Instrument head will not
connect
Spot is not centered
Dim light output
Aperture or lens wheel
will not turn or does so
with difficulty
Visible corrosion on in-
strument
Possible Cause
Aperture dial is between
positions
Lamp Burned out
Wrong Lamp Installed
Lamp not fully inserted
Thread/connection is
damaged
The aperture dial is not
centered
Lamp near end of useful
life
Batteries are depleted
Dirt in or damage to
mechanism
Damage from excessive
moisture
Corrective Action
Rotate Aperture dial
Replace the lamp with ADC part
#5210-4
Replace the lamp with ADC part
#5210-4
Re-insert lamp
Service required
Move aperture dial to the full
detent position
Replace the lamp with ADC
part #5210-4
Replace with fresh “C” batteries
Service required
Service required
Ophthalmoscope Troubleshooting Guide