© Adam Equipment Company 2008
15
7.4
F4
-RE-ZERO RANGE
To enter this parameter, press the
[Tare]
key when “
F4 reZ
” is shown.
The display will show the current re-zero range.
Press the
[Limit]
key to scroll through the options.
Press
[Tare]
to accept the displayed value and return to the parameter or press the
[Zero]
key to return to the parameter without changing the value.
Then press
[Limit]
to advance to the next parameter.
7.5
F5
-SUCCESSIVE TARE
To enter this parameter, press the
[Tare]
key when “
F5 SCS
” is shown.
The display will show if the successive tare is “
tr on
”
or “
tr oFF
”
.
Press the
[Limit]
key to scroll through the options.
Press
[Tare]
to accept the displayed value and return to the parameter or press the
[Zero]
key to return to the parameter without changing the value.
Then press
[Limit]
to advance to the next parameter.
7.6
F6
-A/D COUNT
To enter this parameter, press the
[Tare]
key when “
F6 Cnt
” is shown.
This parameter allows you to view the A/D counts from the internal A/D converter. This can
be an aid to service.
Press the
[Tare]
key to return to the PARAMETER menu.
The actual values are not critical as long as the ADC appears to be running, the values are
stable (±5) and there are approximately 200,000 -300,000 counts span.
7.7
F7
–LOW VOLTAGE DETECT
This parameter allows detection of low voltage when the battery wears down.