© Adam Equipment Company 2008
13
7.1
F1
-CALIBRATION
To enter the calibration parameter, press the
[Tare]
key when “
F1 CAL
” is displayed.
The display will instruct you to remove any weight from the scale, “
UnloAd
”. Press
[Tare]
.
The display will tell you to add the first weight to the scale: “
Ld
1 OO
kg
”
Place the first calibration weight as suggested in the Table below. Wait for stability then
press the
[Tare]
key.
The display will tell you to add the second weight to the scale: “
Ld
2 00
kg
”
Place the second calibration weight as suggested in the Table below. Wait for stability then
press the
[Tare]
key.
The display will show “
PASS
” if the calibration is OK. Or it will show “
FAIL L
” if calibration
could not be completed correctly.
The scale will return to weighing. Remove the calibration weight.
Suggested Calibration Weights:
Calibration weight 1
Calibration Weight 2
WBW 2/5a
1 kg
2 kg
WBW 4/9a
2 kg
4 kg
WBW 8/18a
5 kg
8 kg
WBW 16/35a
5 kg
15 kg
if the scales are weighing in an imperial unit the calibration can be done with pounds
weights. It is suggested weights of approximately
⅓
to 2/3 maximum and near maximum
be used. For example
Calibration weight 1
Calibration Weight 2
WBW 2/5a
2 lb
5 lb
Summary of Contents for WBW 16