![Adam Equipment WBW 16 Service Manual Download Page 16](http://html.mh-extra.com/html/adam-equipment/wbw-16/wbw-16_service-manual_2846146016.webp)
© Adam Equipment Company 2008
14
WBW 4/9a
4 lb
9 lb
WBW 8/18a
10 lb
15 lb
WBW 16/35a
20 lb
30 lb
Other weights can be used as long as the value entered matches the weights used.
7.2
F2
– CAPACITY
To enter this parameter, press the
[Tare]
key when “
F2 CAP
” is shown.
The display will show the current capacity.
Press the
[Limit]
key to scroll through the options.
Press
[Tare]
to accept the displayed value and return to the parameter or press the
[Zero]
key to return to the parameter without changing the value.
Then press
[Limit]
to advance to the next parameter.
NOTE:
The scale must be originally built for the capacity selected. The 2 kg unit uses the
3 kg load cells, 4 kg unit uses 5 kg load cells, the 8 kg unit uses 10 kg load cells and the
16 kg unit uses 20 kg load cells.
7.3
F3
-INITIAL ZERO RANGE
To enter this parameter, press the
[Tare]
key when “
F3 int
” is shown.
The display will show the current initial zero range.
Press the
[Limit]
key to scroll through the options.
Press
[Tare]
to accept the displayed value and return to the parameter or press the
[Zero]
key to return to the parameter without changing the value.
Then press
[Limit]
to advance to the next parameter.