Page 34
This parameter allows you to view the A/D counts from the internal A/D
converter. This can be an aid to service.
Press the [
0/T
] key to return to the PARAMETER menu.
Press the [Print/M+/Esc] key to return to weighing.
Typical value at zero is 30,000-90,000 (approx.).
Typical value at full capacity is 500,000 (approx.).
F8 –ZERO MODE
To enter this parameter, press the [
0/T
] key when “
F8 ZEm
” is shown.
Select the Zero mode desired. In all but special cases Zero Mode 1 is used.
The other 2 zero modes are for unique locations in the world and effect the +/-
range of the zero.
Press the [Pcs/ ] key to change the value.
Press [
0/T
] to accept the displayed value.
Press [Print/M+/Esc] to return to weighing.