10.4.3.3
Dynamic (Animal) Weighing Setup Parameters
When the “
Dynamic
” symbol is displayed, and you have selected “
SEtvP
SEtvP
SEtvP
SEtvP
” to set up the
balance for dynamic weighing (see section 10.4.3.3 on Dynamic Weighing Setup
Parameters):
Press [Setup] to select “
SEtvP
SEtvP
SEtvP
SEtvP
” to change the dynamic weighing mode settings.
The display will show “
MODE
MODE
MODE
MODE
”. Press [Setup] again and use the [Up] or [Down] keys to
select “
AvtO
AvtO
AvtO
AvtO
” or “
mmAnvAL
mAnvAL
mAnvAL
mAnvAL
”.
If “
AvtO
AvtO
AvtO
AvtO
” or “
mmAnvAL
mAnvAL
mAnvAL
mAnvAL
” is selected, the following 4 parameters are available:
•
Threshold
“
THRESH
THRESH
THRESH
THRESH
”
•
Mode
“
mode
mode
mode
mode
”
•
Delay
“
DE
DE
DE
DELAY
LAY
LAY
LAY
”
•
Test time
“
TEST T
TEST T
TEST T
TEST T
”
10.4.3.3.1
Threshold “
THRESH
THRESH
THRESH
THRESH
” (For Auto mode only)
Press [Setup] when “
THRESH
THRESH
THRESH
THRESH
” is shown and the display will next show the minimum
weight of the item required by the balance to start the process for dynamic weighing. The
value shown will be the current value in the last selected unit.
The minimum threshold value can be changed from 1.0 to 100 grams using the keypad
numeric entry method. If a value outside this range is selected then it will not be accepted
and “
ER LOW
ER LOW
ER LOW
ER LOW
” or “
ER HIGH
ER HIGH
ER HIGH
ER HIGH
” will be displayed followed by return to the weight entry screen
again.
To confirm the desired value, press [Setup] or to escape without changing the value,
press [Mode].
10.4.3.3.2
Mode
“
mode
mode
mode
mode
”
Auto “
AvtO
AvtO
AvtO
AvtO
” or Manual “
mmAnvAL
mAnvAL
mAnvAL
mAnvAL
” modes are available. Whichever mode is visible when
[Setup] is pressed becomes the active mode. Auto starts dynamic weighing test as soon
as weight exceeding a set threshold is loaded on the pan. Manual requires the user to
load the pan and then press a button before weighing commences.
10.4.3.3.3
Delay
“
DELAY
DELAY
DELAY
DELAY
”
Press [Setup] when “
DELAY
DELAY
DELAY
DELAY
” is shown and the display will next show the number of
seconds pause before the sampling starts. The Delay value can be changed to between 0-
99 seconds using the keypad numeric entry method. If a value outside this range is
selected then it will not be accepted and “
ER LOW
ER LOW
ER LOW
ER LOW
” or “
ER HIGH
ER HIGH
ER HIGH
ER HIGH
” will be displayed followed
by return to the time entry screen again.
To confirm the desired value, press [Setup] or to escape without changing the value,
press [Mode].
10.4.3.3.4
Test time
“
TEST T
TEST T
TEST T
TEST T
”
Press [Setup] when “
TESt t
TESt t
TESt t
TESt t
” is shown and the display will next show the number of
seconds over which the balance will average to compute the final weight. The Test time