EXTERNAL CALIBRATION (NOT AVAILABLE ON APPROVED MODELS)
1)
Press the
[on/off]
key to turn the power on.
2)
Press the
[Smpl]
and
[Print]
keys at the same time during the self‐checking
test.
3)
The display will show
unload.
Remove any weight from the stainless steel pan.
4)
When the stable indicator shows press the
[‐>0/T<‐]
key.
5)
The display will show the first weight that you can use to calibrate the unit.
You can change this value by pressing the
[Mode]
key. The weights that can
be used are as follows:
Model #
HCB 123
HCB 153
HCB 302
HCB 602
HCB 602H
Weight 1
60g
50g
100g
200g
200g
Weight 2
120g
100g
200g
400g
400g
Weight 3
‐
150g
300g
600g
600g
Model #
HCB 1002 HCB 1502
HCB 2202
HCB 3001
HCB 6001
Weight 1
500g
500g
1000g
1000g
2000g
Weight 2
1000g
1000g
2000g
2000g
4000g
Weight 3
‐
1500g
‐
3000g
6000g
1)
Once the calibration weight has been selected press the
[‐>0/T<‐]
key
2)
The display shows
LoAd.
Place your weight on the pan. Press
[‐>0/T<‐]
key
once the stable sign is shown.
3)
The display will show
PASS.
Remove the weight from the pan.
NOTE:
If the calibration fails retry. The balance will show
FAlL H
(when the weight is
higher) or
FAlL L
(when the weight is lower). Repeat the process using the correct
calibration weight.