8.0
CALIBRATION
8.1
SETTING
CAPACITY
AND
DIVISION
To set up the capacity and division follow the procedure below:
With the indicator turned off, press and hold down the [
PLU]
and
[CE/Backlight]
keys and then turn the scale on using the rocker switch at the right back hand side
of the indicator. The display will show:
Enter 123456 – which is the default passcode and you should see the following
screen:
Press the [
Print/Save]
button, this will take you into the system parameters
(P1
FIL
is to do with filtering) , press the
[Print/Save]
button again, and you should see the
system parameter for settings capacity & divisions (
P2 C-d
), screen should look
similar to as follows: