© Adam Equipment Company 2019
P5 EC
Calibration parameter record. User can review or use it for setting. When
calibration parameters are lost, user can use this record to calibrate without
weight. Note – we would recommend using P4 Cal setting and doing a fresh
calibration as this is the best method
P1 Rul
Calibration parameter
(
the number is in the unit
weight window, this value affects the weighing
accuracy directly. Please note down this parameter
after calibration. When this parameter lost, user can
input the value by number key which been noted to
calibration without weight. Press [M+] key, the digit of
total weight window will keep blinking. Then input the
value and press [PLU] to store and enter to another
parameter.
P2 Ld0
This is the calibration parameter is for multi-range
calibration.
(
If user did not do multi-range calibration;
do not need note this parameter. But need set P3 to
OFF.
)
how to use and set this parameter, it is the
same as standard calibration.
P2 Ld1
The same as Ld0.
P2 Lr0
The same as Ld0.
P2 Lr1
The same as Ld0.